Beckman Coulter Inc. recalls UIBC (Unsaturated Iron Binding Capacity)
Reason for recall
Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
UIBC (Unsaturated Iron Binding Capacity)1 code
- OSR61205
Lot / code information
All Lots/(01)15099590011925
What the firm is doing
On July 10, 2025, Beckman Coulter issued a "Urgent Medical Device Recall" via mail and E-Mail to affected consignees. Beckman Coulter asked consignees to take the following action: 1. Sharing the content of this letter with your laboratory and/or Medical Director to evaluate the requirement for a retrospective review of UIBC results for hemolysed samples. 2. Discontinuance or disposal of this product is not necessary. 3. Per the IFU, avoid hemolysed samples when using the UIBC assay. 4. If the LIH influence check setting for hemolysis has been enabled on AU/DxC AU instruments for UIBC, update the setting per the instructions in the letter. 5. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. 6. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. 7. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email or manually, complete and return the enclosed Response Form.
DistributionShow detailsHide
US: AK, AL, AR, CO, CT, DC, DE, ID, UT OUS: Algeria, Argentina, Australia, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czechia, Egypt, El Salvador, Ethiopia, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Republic of, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Maldives, Mexico, Morocco, Mozambique, Myanmar, Nepal, Netherlands, Nigeria, North Macedonia, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Somalia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Province of China, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Uzbekistan, Viet Nam, Yemen.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2264-2025
- FDA 510(k) clearance · K041793The device's official FDA premarket clearance record
- FDA device classification · JMOOfficial FDA classification for this device type
- CFR regulation · 21 CFR 862.1415The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Beckman Coulter Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm