Class IOngoingZ-2333-2025

Baxter Healthcare Corporation recalls Novum IQ Syringe Pump

Baxter Healthcare CorporationDeerfield, IL, United StatesReported Sep 10, 2025 · 10 months ago
Legal News Analyst ·

Reason for recall

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the Novum IQ Large Volume Pump (LVP) and Novum IQ Syringe Pump (SP) due to software anomalies that may result in a blank Run screen (LVP and SP) and/or false motor movement system error (SP only).

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Novum IQ Syringe Pump, Product Code 40800BAXUS, infusion pump

Lot / code information

UDI/DI 05413765852428, All Serial Numbers

What the firm is doing

Baxter issued an "URGENT MEDICAL DEVICE CORRECTION" to its consignees on 08/04/2025 via USPS first class mail. The notice described the problem, hazard involved, and actions to be Taken by customers: 1. Users should rely on their clinical judgment. When providing patient care involving high risk medications and/or critical illness, users should consider utilizing an alternate pump, if available and appropriate for the patient's treatment. 2. If an alternate pump is not available or appropriate for patient care you may continue to use the Novum IQ LVP and Novum IQ Syringe Pump consistent with the following actions: - If a blank Run screen is displayed during an infusion on the Novum IQ LVP, the user should open the door, press the tubing into load point 3, then load point 4, to turn the light green, and close the door. If a blank Run screen is displayed during an infusion on the Novum IQ Syringe Pump, the user should unload, then reload the syringe. If these actions do not resolve the blank screen, both LVP and Syringe Pump users should power-cycle the pump, select "Yes" to the new patient prompt, and reprogram the infusion. - To prevent a false motor movement system error, do not program the Novum IQ syringe pump to infuse below the Minimum Recommended Flow Rate for a syringe size of 10 mL or greater. 3. Acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. 4. Please forward a copy of this communication to any departments within your institution who use the affected product. 5. Print and post Attachment A in areas frequented by potential users. For general questions regarding this communication, or if you experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867 (select option 2, then option 2 again) Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.

DistributionShow details

US distribution including Puerto Rico and OUS (International) Canada

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2624-2026

Medtronic Heart Valves Division recalls Medtronic Harmony Delivery Catheter System

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a man…

  • Pulmonary Valve Prosthesis Percutaneously Delivered
  • Process control
Medtronic Heart Valves…CAJul 8, 2026
Class IOngoingZ-2622-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026
Class IOngoingZ-2621-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026