PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Best, NetherlandsReported Jun 24, 2026 · 24 days ago
Legal News Analyst ·

Reason for recall

Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Philips Allura Xper with the following Model Numbers: Model # 722001 for Allura Xper FD10C; Model # 722002 for Allura Xper FD10F; Model # 722003 for Allura Xper FD10; Model # 722005 for Allura Xper FD10/10; Model # 722006 for Allura Xper FD20; Model # 722008 for Allura Xper FD20 Biplane; Model # 722010 for Allura Xper FD10; Model # 722011 for Allura Xper FD10/10; Model # 722012 for Allura Xper FD20; Model # 722013 for Allura Xper FD20 Biplane; Model # 722015 for Allura Xper FD20 OR Table; Model # 722019 for Allura Xper FD10/10 OR Table; Model # 722020 for Allura Xper FD20 Biplane OR Table; Model # 722022 for Allura Xper FD10 OR Table; Model # 722023 for Allura Xper FD20 OR Table; Model # 722025 for Allura Xper FD20 Biplane OR Table; Model # 722026 for Allura Xper FD10; Model # 722027 for Allura Xper FD10/10; Model # 722028 for Allura Xper FD20; Model # 722029 for Allura Xper FD20/10; Model #722033 for Allura Xper FD10 OR Table; Model #722035 for Allura Xper FD20 OR Table; Model #722038 for Allura Xper FD20/20; Model #722039 for Allura Xper FD20/20 OR Table; Model #722058 for Allura Xper FD20/15; Model #722059 for Allura Xper FD20/15 OR Table.

Lot / code information

All codes

What the firm is doing

Philips will send a customer notification dated May 18, 2026 to the users of the affected systems to alert them of the issue and provide the actions that should be taken to prevent risks to patients Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit their site and replace the hand switch with a new design hand switch.

DistributionShow details

US and ROW.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.