Class IOngoingZ-2430-2025

Boston Scientific Corporation recalls WATCHMAN FXD Curve Access Sys Dbl

Boston Scientific CorporationMaple Grove, MN, United StatesReported Sep 10, 2025 · 10 months ago
Legal News Analyst ·

Reason for recall

Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

WATCHMAN FXD Curve Access Sys Dbl, OUS, Material Number (UPN) M635TS80020; intended to provide vascular and transseptal access for the family of WATCHMAN FLX LAAC Devices with Delivery Systems.

Lot / code information

GTIN
00191506013837, ALL NON-EXPIRED BATCHES

What the firm is doing

Boston Scientific issued an Important Medical Device Advisory notice to its consignees on 07/29/2025 via FedEx. The notice explained the problem with the device, procedures under which the likelihood of the problem occurs, and risk to the patient. The WATCHMAN Access Systems Instructions for Use (IFUs) and WATCHMAN physician training will be updated to emphasize instructions related to Access System air management. This update will strengthen the information provided to clinicians regarding the potential for air embolism during WATCHMAN procedures performed under conscious or deep sedation and provide potential mitigation strategies. No devices are being removed. Instructions: 1. Review IFU updates related to air embolism as detailed in Appendix 2. These updates will be added to the WATCHMAN Access Systems IFUs and WATCHMAN Physician Training. 2. Forward this letter to any other clinicians in your medical facility who perform WATCHMAN procedures and to any facilities where affected devices have been transferred, including hospitals or sites within your network. If you are a distributor, this notice must be forwarded to your customers to ensure notification of this Medical Device Advisory is carried out to the end-user level. 3. Complete and return the enclosed Acknowledgment Form. If additional assistance is required or more information regarding this communication, customer should contact your local Boston Scientific representative or Sr. Quality Systems Manager at 763-494-1133.

DistributionShow details

Worldwide.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2624-2026

Medtronic Heart Valves Division recalls Medtronic Harmony Delivery Catheter System

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a man…

  • Pulmonary Valve Prosthesis Percutaneously Delivered
  • Process control
Medtronic Heart Valves…CAJul 8, 2026
Class IOngoingZ-2622-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026
Class IOngoingZ-2621-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026