Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-2437-2026

Medline Industries, LP recalls Centurion convenience kits: CENTRAL LINE DRESSING SV43 FCP DYNDC2356B DIALYSIS C…

Medline Industries, LPNorthfield, IL, United StatesReported Jun 24, 2026 · 24 days ago
Legal News Analyst ·

Reason for recall

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Centurion convenience kits: CENTRAL LINE DRESSING SV43 FCP DYNDC2356B DIALYSIS CATHETER MAINTENANCE KIT DT17570 DIALYSIS DRESSING CHANGE KIT DT21370B DRESSING CHANGE TRAY - GOLD DT21115B PORT DRESSING CHANGE KITDYNDC2815A
    UDI-DI 10653160327271
    13 affected lots
    2025102090DT17570106531602668912025102290DT21370B106531603606432025102290DT21115B
    +5 more106531603606812025111490DYNDC2815A106531603570702025110590

What the firm is doing

Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.

DistributionShow details

US Nationwide distribution. OUS distribution pending.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls