Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-2439-2026

Medline Industries, LP recalls Medline convenience kits: 20CM CVC INSERTION KIT ECVC6805B 2L 8.5FR 16CM CVC INS…

Medline Industries, LPNorthfield, IL, United StatesReported Jun 24, 2026 · 24 days ago
Legal News Analyst ·

Reason for recall

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Medline convenience kits: 20CM CVC INSERTION KIT ECVC6805B 2L 8.5FR 16CM CVC INSERTION KIT ECVC8055A VANTEX PI CVC 7F, 3L, 20CM BUNDLE ECVC6915AECVC6985B
    UDI-DI 10653160387428
    10 affected lots
    2025120890ECVC8055A106531603878482025112690ECVC6915A106531603894392025102790ECVC6985B
    +2 more106531603882722025103190

What the firm is doing

Firm began notifying consignees on 4/27/2026. Customers were instructed to request stickers to over-label affected convenience kits with a warning label. Warning label indicates for user to remove the affected component and replace it with product from supply. No kits are to be returned.

DistributionShow details

US Nationwide distribution. OUS distribution pending.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.