Skip to content
RecallWatchMedical Device Safety
Class IOngoingZ-2454-2025

Mo-Vis BVBA recalls Micro Joystick R-net. Electrical wheelchair component.

mo-Vis BVBANevele, BelgiumReported Sep 10, 2025 · 10 months ago
Legal News Analyst ·

Reason for recall

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Micro Joystick R-net. Electrical wheelchair component.

Lot / code information

Model
P002-52
UDI
05407008320027
Serial #
1000 to 1690

What the firm is doing

An URGENT: MEDICAL DEVICE RECALL R-NET JOYSTICKS notification dated 07/10/2025 was sent to consignees instructing them to upgrade affected devices according to the provided Annex 2 or by emailing contact@mo-vis.com to coordinate return. The software upgrade of affected devices should be conducted during the next scheduled annual maintenance of the joystick. These actions should be verified. Questions can be directed via email to contact@mo-vis.com or by phone at +32 9 335 28 60.

DistributionShow details

US Nationwide distribution in the state of TX.

Class IWhat this means

A reasonable probability that use of the product will cause serious adverse health consequences or death.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class I recalls

Class IOngoingZ-2624-2026

Medtronic Heart Valves Division recalls Medtronic Harmony Delivery Catheter System

Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a man…

  • Pulmonary Valve Prosthesis Percutaneously Delivered
  • Process control
Medtronic Heart Valves…CAJul 8, 2026
Class IOngoingZ-2622-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026
Class IOngoingZ-2621-2026

Windstone Medical Packaging, Inc. recalls A M S

Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.

  • General Surgery Tray
  • Material/Component Contamination
Windstone Medical Pack…MTJul 8, 2026