WASATCH PHOTONICS recalls Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers

WASATCH PHOTONICSMorrisville, NC, United StatesReported Sep 17, 2025 · 10 months ago
Legal News Analyst ·

Reason for recall

Two new potential laser safety failure modes have been identified in the product design. Overcurrent of laser diode, and a fail-unsafe mode. The failure modes have never been found to occur in these models, whether in the field or factory, but could potentially occur under excessive force or extreme, continuous vibration.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Wasatch Photonics 785/830-SR and 785-ER-ILC Raman Spectrometers, Product Models: 785-SR-ILP, 785-SR-ILC, 830-SR-ILP, 830-SR-ILC785-ER-ILC
    Model numbers.

What the firm is doing

Notifications will be sent out to all purchasers of affected models. Notification will include an offer of free RMA (return to factory for repair, then shipment back to customer). All repairs will be performed at the site of original manufacture in Morrisville, NC. Customer notifications to be delivered by 31-Aug-2025.

DistributionShow details

U.S.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.