Integra LifeSciences Corp. (NeuroSciences) recalls Brand Name: MediHoney ¿ Product Name: MediHoney¿ Paste Model/Catalog Number: 31505 P…

Integra LifeSciences Corp. (NeuroSciences)Princeton, NJ, United StatesReported Sep 10, 2025 · 10 months ago
Legal News Analyst ·

Reason for recall

Packaging failures were identified which could lead to a breach in the sterile barrier.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Brand Name: MediHoney ¿ Product Name: MediHoney¿ Paste Model/Catalog Number: 31505 Product Description: MediHoney¿ Paste with Active Leptospermum Honey is a moist dressing made of 100% Active Leptospermum medical grade Honey. The dressing helps maintain a moist environment conducive to wound healing and helps aid and support autolytic debridement. The dressing can be removed without causing trauma to the wound when a primary dressing is used.

Lot / code information

Catalog #
31505
UDI
10381780486930
Lot #
s: 2150, 2336, 2151, 2337, 2201, 2414, 2209, 2415, 2217, 2417, 2232, 2428, 2241, 2434, 2301, 2435, 2312, 2505, 2313, 2508, 2314, 2509, 2328

What the firm is doing

On August 5, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions for Customers: If you have units of the impacted product remove them immediately from service and quarantine them. Forward this notice to those who utilize the product so they are aware of this recall and can identify any affected product that may remain in clinical areas. Complete the entirety of the Acknowledgement Form and send via email to FCA4@integralife.com or FAX to 1-609-750-4220. Keep a copy of the form for your records. Actions for Distributors: If you have products listed in Table 1, remove the product from further distribution. Complete the entirety of the Acknowledgement Form and send via email to FCA4@integralife.com or FAX to 1-609-750-4220. Keep a copy of the form for your records. Check your customer traceability records and notify them if they have any shipments of above catalog and lot numbers. Modify the acknowledgement form to create one from you to your customers. Collect completed response forms and affected product from your customers and indicate total quantities and lots in distributor reply form. For questions regarding these instructions, please contact Customer Service: Monday - Friday 8:00 AM 8:00 PM EST USA: 1-800-654-2873 email: custsvcnj@integralife.com Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Integra is dedicated to patient safety and manufacturing excellence. We continue to make numerous quality improvements and investments in our facilities around the world. In addition, we are improving our processes in accordance with applicable regulations. We sincerely apologize for any inconvenience this voluntary recall may cause and thank you for your cooperation in this effort.

DistributionShow details

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and the countries of Canada, Australia, New Zealand, Korea, Guam, Malaysia.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.