Class IIOngoingZ-2523-2025

ICU Medical Inc. recalls Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Non-vente…

ICU Medical Inc.Minneapolis, MN, United StatesReported Sep 10, 2025 · 10 months ago
Legal News Analyst ·

Reason for recall

Certain CADD 50 mL and 100 mL Medication Cassette Reservoirs may exhibit weakened medication bag welds along the bag perimeter inside the plastic housing. This could result in medication leakage along the bag perimeter.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Smiths Medical CADD Medication Cassette Reservoir with clamp and female Luer. Non-vented stopper included. 100mL, Product Code21-7002-24
    18 affected lots
    60220106026827602682860268296026830602683160377506037751
    +10 more6037752603775360539806053981605398260539836062689606268860638456062691

What the firm is doing

ICU Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 08/06/2025 via registered mail and email. The notice explained the issue, potential risk can requested the following: Customer Required Actions: Please complete the following actions below. 1. Review your inventory of CADD Medication Cassette Reservoirs to confirm if any of the affected lots are within your possession. 2. Quarantine the affected product and destroy or discard it immediately following your institution s process for destruction or discarding. 3. Return the attached Customer Response Form to ICUMedical8176@sedgwick.com within ten (10) days of receipt to acknowledge your understanding of this notification. 4. DISTRIBUTORS: Immediately forward this notice to all users to whom you distributed these products. Request that they complete the Customer Response Form and return it to ICUMedical8176@sedgwick.com. For further inquires: - To report adverse events or product complaints: Global Complaint Management: productcomplaints@icumed.com; 1-(866)-216-8806 - Additional information or to request replacement product: Customer Service: customerservice@icumed.com; 1-(800)-258-5361 - Questions about this communication: Field Action Processing: ICUMedical8176@sedgwick.com; 1-(844)-861-6220

DistributionShow details

Worldwide distribution.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.