Boston Scientific Neuromodulation Corporation recalls Vercise Deep Brain Stimulation Systems Surgical Implant Manual
Reason for recall
Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Vercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number:92328632-09UDI-DI 08714729985044
What the firm is doing
On 7/8/2025, correction notices were mailed in the U.S. and mailed, emailed, or hand delivered outside the U.S. Healthcare Professionals were asked to do the following: 1) Review the Surgical Implant Manual updates. 2) If you are a facility that has sent products to another hospital or a facility within your network, ensure this notification is forwarded to them. 3) A patient letter is enclosed with this communication, which can be shared with your patient and/or included within the patient's medical record. 4) Monitor per relevant IFU recommendations for any clinical observations of high monopolar impedances, undesired sensation, sudden loss of therapy, return of pre-implant symptoms and/or Bluetooth connectivity challenges, as these may be signals of potential feedthrough wire break(s) (component damage). 5) If you are a distributor, this notification must be forwarded to your customers to ensure this notification is carried out to the end-user level. 6) To provide awareness of this information, share this letter with any other clinicians in your hospital who use the system. 7) Maintain a copy of this letter in your facility's records. 8) Complete and return the acknowledgement form via email to BSCFieldActionCenter@bsci.com report all device-related incidents or quality concerns experienced with the use of these devices to the firm at BSN.ComplaintCallCenter@bsci.com
DistributionShow detailsHide
Worldwide - US Nationwide distribution including in the states of MN, FL, CO, MD, NY, NE, PA, WI, AZ, TN, MO, TX, MI, MT, MA, WV, AL, VA, NC, CA, OR, WA, LA, OH, NJ, NH, SD, GA, CT, NV, UT, IN, ID, IL, ME, SC, DC, OK, KY, AK, HI, RI, AR, KS, NM and the countries of Argentina, Brazil, Chile, Ecuador, Colombia , Mexico, Canada, Japan, Algeria, Austria, Azerbaijan, Belgium, Bulgaria, Croatia, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Iran, Ireland, Israel, Italy, Jordan, Kazakhstan, Kuwait, Lebanon, Libya, Luxembourg, Malta, Morocco, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkiye, UAE, Ukraine, United Kingdom, Hong Kong, China, Singapore, Korea, Taiwan , India, Thailand.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2558-2025
- FDA device classification · NHLOfficial FDA classification for this device type
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Boston Scientific Neuromodulation CorporationSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm