Dexcom, Inc. recalls Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US…
Reason for recall
Certain lots of Dexcom G7 sensors originally designated as scrap and intended for destruction were stolen during the destruction process. The sensors were later sold to pharmacies and distributors by Pharmsource, LLC. Lot 1725204004 were designated for destruction because the G7 sensors were not properly sterilized and have an increased risk of skin infection if used. Lot 1725069002 were designated for destruction due to an increased risk of having no sensor readings available or experiencing sensor inaccuracy that includes the potential for missed detection of a hyperglycemic event or incorrect treatment decisions that can result in patient injury, including medical intervention to prevent permanent injury or impairment. Additionally, due to the unknown storage and handling and storage conditions these sensors may have undergone during the distribution, these sensors may present additional risk.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Brand Name: Dexcom G7 Continuous Glucose Monitoring System Product Name: Dexcom G7 US Retail Sensor/Transmitter 1-pack Model/Catalog Number: STP-AT-012 Software Version: N/A Product Description: Dexcom G7 US Retail Sensor/Transmitter 1-pack Component: Dexcom G7 Continuous Glucose Monitoring SystemUDI-DI 003862700041093 affected lots0038627000410917252040041725069002
What the firm is doing
On 05/26-27/2026, the firm issued a Press Release and sent an "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION" informing customers and the public that Dexcom is aware that two lots (1725204004 and 1725069002) of Dexcom G7 sensors, originally designated as scrap and intended for destruction, were stolen and distributed. For Patients/Healthcare Providers: 1. Check lot number on G7 sensor outer box and compare with the listed lot numbers in this notification. If the lot number is listed, the device is affected. If affected, skip to step 3 below to document affected serial number(s). 2. Determine if current or unused sensor(s) is affected based on serial number by visiting www.dexcom.com/theft-check and enter serial number into the serial number look-up. Reference https://www.dexcom.com/faqs/how-to-find-dexcom-g7-serial-number for instructions. 3. If you have affected product(s), immediately discontinue use and dispose of them. Contact Dexcom Technical Support at 1-844-478-1600 with questions or to request a no-cost replacement. Important: Keep record of affected serial number for traceability purposes This is required to confirm no cost replacement. 4. Replace the affected sensor with an unaffected sensor for glucose monitoring. If you do not have an unaffected G7 sensor, use a blood glucose meter for treatment decisions until you receive a replacement sensor. 5. Contact healthcare provider if you have symptoms of infection or inflammation redness, swelling, or pain at the insertion site. Independent Pharmacies: Check inventory if purchased G7 sensors from Pharmsource, LLC. If you have product from the lots listed, immediately dispose of the product. Contact Dexcom with any related inquiries at inquiries@dexcom.com.
DistributionShow detailsHide
Dexcom does not have distribution records for this product due to its unauthorized release. However Dexcom is able to identify sensor session starts for the affected lots through the Dexcom G7 CGM App user performance data. Based on App user performance data, Lot 1725204004 has sensor session starts in the United States. Based on App user performance data, Lot 1725069002 has sensor session starts in the United States and the countries of Australia, Austria, Canada, Czech Republic, Finland, France, Germany, Hong Kong, Israel, Italy, Japan, South Korea, Kuwait, Netherlands, Poland, Portugal, Qatar, Singapore, Spain, Switzerland, Turkey, United Arab Emirates,and the United Kingdom.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2574-2026
- FDA 510(k) clearance · K240902The device's official FDA premarket clearance record
- FDA device classification · QBJOfficial FDA classification for this device type
- CFR regulation · 21 CFR 862.1355The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Dexcom, Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems S…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
Zimmer Surgical Inc recalls Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S¿ 4000 TS Tourniquet Systems D…
Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find anot…
- Pneumatic Tourniquet
- Under Investigation by firm
W L Gore & Associates, Inc. recalls GORE¿ TAG¿ Thoracic Branch Endoprosthesis Side Branch Component Catalog Numbers (by Re…
Due to catheter separation
- Aortic Aneurysm Treatment Endovascular Graft System
- Under Investigation by firm