Class IIOngoingZ-2581-2026

Medline Industries, LP recalls Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGI…

Medline Industries, LPNorthfield, IL, United StatesReported Jul 1, 2026 · 17 days ago
Legal News Analyst ·

Reason for recall

Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIOGRAM PACK-LF DYNJ21574J JUDKINS PACKDYNJ51126
    UDI-DI 10889942153497
    10 affected lots
    21ALA58621CLA39821ELA40821FLA14421FLA78821HLA42021ILA431DYNJ51126
    +2 more1088994212282021ADA917

What the firm is doing

This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.

DistributionShow details

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.