Philips North America LLC recalls Philips CT systems labeled as the following: 1. Ingenuity CT Family Brazil
Reason for recall
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Philips CT systems labeled as the following: 1. Ingenuity CT Family Brazil, Model Number: 728325.
Lot / code information
- Model
- 728325
- UDI
- N/A
- Serial #
- 338014, 338028, 338019
What the firm is doing
Philips Medical notified consignees on about 08/11/2025 via letter. Consignees were instructed that they could continue using the affected systems accordance with the intended use and by following the recommendation: Observe the warning listed in Instructions for Use (IFU) and make sure that the patient is strapped securely to avoid dangling of the hands. Ensure that the patient is placed securely on the patient table and is not in danger of falling. Additionally, consignees were instructed to complete and return the provided response form, to circulate Urgent Medical Device Correction Letter to all users of this device so that they are aware of the issue and retain this letter with your system(s) until your system is inspected and corrected if necessary. Philips will contact consignees to schedule a time for a Philips Field Service Engineer (FSE) to visit each site and to inspect the affected system (FCO 72800832) and correct the system if necessary (FCO 72800833).
DistributionShow detailsHide
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Mauritius, Mexico, Netherlands, Norway, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, T¿rkiye, United Kingdom, Utd.Arab.Emir., Vietnam.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2603-2025
- FDA 510(k) clearance · K033326The device's official FDA premarket clearance record
- FDA device classification · JAKOfficial FDA classification for this device type
- CFR regulation · 21 CFR 892.1750The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Philips North America LlcSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) BASIC PACK
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Eye Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) KIT
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Orthopedic Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) CHI STD NEURO CRANI OPT 2
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Neurological Tray
- Nonconforming Material/Component