Medline Industries, LP recalls Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIA…
Reason for recall
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL LINE TOTEDYNJ8407UDI-DI 10198459704543Affected lot26DBB599
What the firm is doing
On May 28, 2026, the firm notified customers via letters titled "URGENT MEDICAL DEVICE CORRECTION." Customers were referred to BD's recall notice, which provides advice for users and instructions to mitigate hub breakage issues. Product return is not required. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
DistributionShow detailsHide
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2608-2026
- FDA device classification · OFFOfficial FDA classification for this device type
- CFR regulation · 21 CFR 870.1140The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Medline Industries, LPSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) BASIC PACK
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Eye Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) KIT
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Orthopedic Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) CHI STD NEURO CRANI OPT 2
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Neurological Tray
- Nonconforming Material/Component