Medline Industries, LP recalls Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardia…
Reason for recall
Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Reprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac stimulation during diagnostic electrophysiologic studies; Medline Item Numbers 81102RH 81104RH 81107RH 81174RH 81402RH 81404RH 81405RH 81472RH 81473RH 81474RH 81531RH 81532RH
Lot / code information
- UDI
81102RH 10197344133239; 81104RH 10197344133246; 81107RH 10197344133260; 81174RH 10197344133291; 81402RH 10197344133352; 81404RH 10197344133376; 81405RH 10197344133383; 81472RH 10197344133413; 81473RH…Show all
81102RH 10197344133239; 81104RH 10197344133246; 81107RH 10197344133260; 81174RH 10197344133291; 81402RH 10197344133352; 81404RH 10197344133376; 81405RH 10197344133383; 81472RH 10197344133413; 81473RH 10197344133420; 81474RH 10197344133437; 81531RH 10197344133512; 81532RH 10197344133529; LOTS EP250212 EP250214
What the firm is doing
On July 31, 2025, the firm notified customers via first class mail and email of the potential product issue. An updated letter was sent to customers on 8/13/25 with additional item and lot numbers. Affected products are to be returned for credit. Customers were instructed to immediately identify and quarantine all affected product within their possession. Once they have completed and returned the response form listing their affected inventory, they will receive return labels and RGA information. Customers will receive credit when the returned product is received.
DistributionShow detailsHide
Domestic US distribution nationwide. International distribution pending. No international distribution.
A reasonable probability that use of the product will cause serious adverse health consequences or death.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2612-2025
- FDA 510(k) clearance · K242225The device's official FDA premarket clearance record
- FDA device classification · NLHOfficial FDA classification for this device type
- CFR regulation · 21 CFR 870.1220The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Medline Industries, LPSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class I recalls
Medtronic Heart Valves Division recalls Medtronic Harmony Delivery Catheter System
Devices containing inner shaft assemblies from specific manufacturing lots may have an increased potential for distal tip detachment from the delivery system under certain procedural or anatomical conditions due to a man…
- Pulmonary Valve Prosthesis Percutaneously Delivered
- Process control
Windstone Medical Packaging, Inc. recalls A M S
Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
- General Surgery Tray
- Material/Component Contamination
Windstone Medical Packaging, Inc. recalls A M S
Kits contain Webcol Large Alcohol Prep Pad which were recalled by Cardinal Health following the discovery of a contaminant (Paenibacillus phoenicis) during a routine sterilization dose audit.
- General Surgery Tray
- Material/Component Contamination
