Class IIOngoingZ-2663-2026

Bard Access Systems, Inc. recalls CK000980A PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236150 P425410…

Bard Access Systems, Inc.Salt Lake City, UT, United StatesReported Jul 8, 2026 · 10 days ago
Legal News Analyst ·

Reason for recall

Due to lidocaine ampoules being manufactured under deficient manufacturing practices by manufacturer, lidocaine ampoules are currently being recalled by supplier.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • CK000980A PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741236150 P4254108 PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741129278 P4254108D PowerMidline Catheter 4 Fr Dual-Lumen UDI-DI Code: 00801741129353 The PowerMidline" Catheter is intended for short term peripheral access for selected intravenous therapies, blood sampling, and power injection of contrast media
    UDI-DI Code
    7 affected lots
    REKX0112P425410800801741129278REKW0949P4254108D00801741129353REJZ1337

What the firm is doing

On 05/21/2026, the firm FedEx'd an "URGENT: Medical Device Recall" letter informing customers/distributed that BD has become aware that the FDA has placed Lidocaine manufactured by Huons Co., Ltd. on import alert #66-40 - Detention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPs. BD is recalling products that contain the affected lidocaine ampules. Customers are instructed to: 1. Not to use the affected Lidocaine ampules. 2. Secure an alternative local Lidocaine drug prior to entering the procedural environment. 3. Upon opening the kit, remove and destroy the drug product from the BD convenience kits/procedure trays at the point of use, within a controlled and sterile environment to maintain the sterility of the remaining components within the kit. All other kit/tray components are safe to use. 4. If affected Lidocaine ampule was previously used without issue, no additional treatment or clinical follow up is required. 5. If affected Lidocaine ampule was previously used with issue, report the incident per institutional guidelines and take any necessary actions. Additional actions to take: 1. Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. 2. Post this notice on all inventory storage locations and any location where the product may be used. 3. Label affected kits/trays using the template provided in Appendix 2 to facilitate easy identification of products containing the affected ampules. The label can be printed using Avery template 5160. a. Apply the labels to the product without covering or obscuring any regulatory or traceability information. b. Apply to the top web of the kit/tray pressing firmly to ensure adhesion, while taking care not to damage or compromise the sterile barrier. For Questions - Contact North American Regional Complaint Center at 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompt

DistributionShow details

Worldwide U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The country of Belgium.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.