Class IIOngoingZ-2711-2026

Elekta, Inc. recalls Brand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator…

Elekta, Inc.Atlanta, GA, United StatesReported Jul 15, 2026 · 7 days ago
Legal News Analyst ·

Reason for recall

Due to safety concerns regarding the detachment of the tip of the intrauterine tube (IU tube) that may remain within a patient.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Brand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: The intended use of the Advanced Gynecological Applicator is intracavitary brachytherapy for cancer treatment of the cervix and endometrium. Optional needles can be placed for interstitial brachytherapy
    11 affected lots
    0871721305500108717213055018087172130550250871721305504908717213055056087172130550630871721305508708717213055094
    +3 more087172130551000871721305511708717213055131

What the firm is doing

On 04/03/2026, an "IMPORTANT FIELD SAFETY NOTIFICATION/URGENT: MEDICAL DEVICE CORRECTION" Letter was sent via email informing customers of a safety concern regarding detachment of the tip of an Intrauterine tube (IU tube) with a 4 mm diameter. Customers are instructed to: To prevent any potential health risks related to a detached tip, we emphasize the importance of adhering to the Instructions for Use: " Do not use an IU tube beyond its 3-year life expectancy or 300 reprocessing cycles, whichever comes first. " During the treatment preparation phase, before insertion of the applicator: Visually inspect parts for indication of damage, weakening, bending, cracks, corrosion, or general deterioration. " Inspect joints and connections on play, strength or other forms of degradation. In addition, we recommend the following: " Inspect the end of the IU tube to ensure that there is no gap between the tip and the fiber tube and that the tip is not detached. " During inspection, insertion and removal: avoid the use of excessive force that could damage the IU tube. " After removal: Ensure that all applicator parts have been removed from the patient and that the tip is not detached. In case you find a damaged or detached IU tube tip, remove it from clinical use and contact Elekta for a replacement. For Questions- Region America Application.Support.NA@elekta.com Tel: +1 855 6935358 https://www.elekta.com Region Asia Pacific HK.TW.Support@elekta.com Tel: + 852 2891 2208 https://www.elekta.com Japan Japan-fco@elekta.com Tel: + 81 3 6722 3800 Fax: +81 3 6436 4231 https://www.elekta.com Region China FCO.CN@elekta.com Tel : + 86 800 810 2550 https://www.elekta.com Region Europe Support.europe@elekta.com Tel : + 46 8 587 254 00 https://www.elekta.com Region Turkey, India & Middle East support.rma@elekta.com Turkey Tel: +90 216 444 6374 India Tel: +1-800-103-7454 Middle East Tel: +00 800 4000 5000 https://www.elekta.com

DistributionShow details

Worldwide - U.S. Nationwide distribution including in the states of AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MO, MS, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TX, UT, VA, WI, and WV. The countries of Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Botswana, Brazil, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iran (Islam Republic of), Iraq, Ireland, Italy, Japan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Rwanda, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Togo, Turkey, Ukraine, United Arab Emirates, United Kingdom, Venezuela, Viet Nam, and Zambia.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-2715-2026

Biomerieux Inc recalls VITEK 2 AST-N447 TEST KIT Product reference 424620

Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.

  • Short Incubation Antimicrobial Susceptibility Automated Test System
  • Process change control
Biomerieux IncMOJul 15, 2026