WALLAC OY recalls GSP Neonatal GALT Kit. Product Codes: 3303-001U
Reason for recall
Reduction in the on-board stability of the GSP GALT assay. This may increase the likelihood of run-QC failures and consequently delay the reporting of patient results.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
GSP Neonatal GALT Kit. Product Codes: 3303-001U, 3303-0010. in vitro diagnostic. Used in FLUORESCENT PROC. (QUAL), GALACTOSE-1-PHOSPHATE URIDYL TRANSFERASE.
Lot / code information
- UDI
- 06438147321704. UDI-PI (PACKING
- Lot #
): (01)06438147321704(17)260531(10)763638 (1076363801); (01)06438147321704(17)260531(10)764085 (1076408501); (01)06438147321704(17)260531(10)765033 (1076503301); (01)06438147321704(17)260731(10)765231…Show all
): (01)06438147321704(17)260531(10)763638 (1076363801); (01)06438147321704(17)260531(10)764085 (1076408501); (01)06438147321704(17)260531(10)765033 (1076503301); (01)06438147321704(17)260731(10)765231 (1076523101); (01)06438147321704(17)260731(10)765316 (1076531601); (01)06438147321704(17)261031(10)766420 (1076642001); (01)06438147321704(17)261031(10)766894 (1076689401); (01)06438147321704(17)261130(10)767873 (1076763801); (01)06438147321704(17)261130(10)768361 (1076836101); (01)06438147321704(17)261130(10)768703 (1076870301); (01)06438147321704(17)270228(10)769355 (1076935501); (01)06438147321704(17)270228(10)769542 (1076954201); (01)06438147321704(17)270228(10)769632 (1076963201); (01)06438147321704(17)270228(10)769671 (1076967101); (01)06438147321704(17)270228(10)769782 (1076978201); (01)06438147321704(17)270228(10)769871 (1076987101); (01)06438147321704(17)270228(10)769975 (1076997501); (01)06438147321704(17)270228(10)770078 (1077007801); (01)06438147321704(17)270228(10)770086 (1077008601); (01)06438147321704(17)270228(10)770351 (1077035101). 2. Product Code: 3303-0010- UDI
- 06438147311347. UDI-PI (PACKING
- Lot #
): (01)06438147311347(17)260531(10)763396 (1076339601); (01)06438147311347(17)260531(10)763396 (1076339602); (01)06438147311347(17)260531(10)764209 (1076420901); (01)06438147311347(17)260531(10)764…Show all
): (01)06438147311347(17)260531(10)763396 (1076339601); (01)06438147311347(17)260531(10)763396 (1076339602); (01)06438147311347(17)260531(10)764209 (1076420901); (01)06438147311347(17)260531(10)764209 (1076420902); (01)06438147311347(17)260731(10)765317 (1076531702); (01)06438147311347(17)260731(10)7653177 (1076531703); (01)06438147311347(17)261031(10)766378 (1076637801); (01)06438147311347(17)261031(10)766943 (1076694301); (01)06438147311347(17)261031(10)767323 (1076732301); (01)06438147311347(17)261130(10)767955 (1076795501); (01)06438147311347(17)261031(10)767956 (1076795601); (01)06438147311347(17)261130(10)769214 (1076921401); (01)06438147311347(17)261231(10)769967 (1076996701)
What the firm is doing
Wallac Oy, a Revvity company, notified consignees on about 05/18/2026 via emailed "URGENT: MEDCIAL DEVICE RECALL" letter. Consignees were instructed to inspect inventory for the affected lots and dispose of affected lots according to local requirements. If consignees do not have another GSP Neonatal GALT kit lot at hand, the newborn screening for classical galactosemia may be continued, with following instructions: - Reconstitute GALT Substrate Reagent by adding 2.8 mL of GALT Assay Buffer and invert gently for 20-30 minutes (note: slightly extended dissolving time). - Reagent cassettes should preferably be used within 12 hours (note: reduced on-board storage time). - Load reagent cassette to GSP without additional delay after reconstitution of Substrate Reagent (one reagent cassette is sufficient for measurement of four full plates (384 wells). - If possible, prioritize GALT plates first if multiple analytes/kits are run within the same instrument. Avoid exceeding the on-board storage time of 12 hours for the reagent cassette. - Run calibration curve on each plate and evaluate the results against the plate-specific curve (note: do not use stored calibration curve). These precautions are aimed at minimizing the fluctuation of the GALT activity results between assays/plates. Consignees were also requested to complete and return the response form provided. Revvity will replace the quantity of excessive kit usage and the affected GSP Neonatal GALT kits disposed from inventory upon receipt of the response form.
DistributionShow detailsHide
Worldwide - US Nationwide distribution in the states and territories of AL, AR, AZ, CO, FL, GA, IL, IN, KY, LA, MA, MD, MN, MO, NJ, OR, PR, SC, TN, TX, VA, WV and the countries of Austria, Belgium, Egypt, France, Germany, Hong Kong, Iraq, Italy, Netherlands, Oman, Philippines, Saudi Arabia, Singapore, Spain, Thailand, United Arab Emirates, Vietnam.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2713-2026
- FDA 510(k) clearance · K100101The device's official FDA premarket clearance record
- FDA device classification · KQPOfficial FDA classification for this device type
- CFR regulation · 21 CFR 862.1315The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find WALLAC OYSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
CMR SURGICAL LIMITED recalls Brand Name: Versius Surgical System US Product Name: Vesius Surgeon Console US Model/…
Secure Boot was mistakenly not enabled at manufacturing time, which presents a potential cybersecurity risk.
- Modular Electromechanical Surgical System
- Software design (manufacturing process)
Think Surgical, Inc. recalls TMINI Femur Tracking Array REF 109003 UDI-DI code: M9461090030 TMINI Tibia Tracking…
Due to manufacturing/design issue where the knob on the tracking array may loosen causing potential misalignment of the array.
- Orthopedic Stereotaxic Instrument
- Process control
Biomerieux Inc recalls VITEK 2 AST-N447 TEST KIT Product reference 424620
Potential for false resistant carbapenem and/or cephalosporin results when analyzing samples with kits from affected lot that have pouch foil fractures.
- Short Incubation Antimicrobial Susceptibility Automated Test System
- Process change control