Class IIOngoingZ-2719-2026

Olympus Corporation Of The Americas recalls Olympus Toomey Evacuator 50cc with Adaptor. Model Number: 188A.

Olympus Corporation of the AmericasCenter Valley, PA, United StatesReported Jul 22, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Olympus Toomey Evacuator 50cc with Adaptor. Model Number:188A
    UDI-DI 00821925001077.Model Number

What the firm is doing

Olympus notified consignees on about 06/05/2026 via Courier, 2-day delivery. Consignees were requested to examine inventory for affected units. If affected units are on hand, replace any existing sterilization instructions for these products with those provided with the notification and add these instructions to any existing copies of the IFU. Consignees were also instructed to acknowledge receipt of the customer letter through the web portal. If affected units were further distributed, Olympus instructed consignees to identify those customers and forward the notification to them. An updated notice was sent to consignees on 06/30/2026 with further clarification regarding instructions to be used to sterilize affected devices going forward.

DistributionShow details

Worldwide distribution - US Nationwide and the countries of Canada, Japan, EU, and Latin America.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-2746-2026

Tosoh Bioscience Inc recalls ST AIA-Pack Testosterone

The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone resul…

  • Testosterones And Dihydrotestosterone Radioimmunoassay
  • Component design/selection
Tosoh Bioscience IncOHJul 22, 2026
Class IIOngoingZ-2744-2026

Daavlin Distributing Company recalls Photera 600 3-D

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

  • Dermatological Ultraviolet Light
  • Software Design Change
Daavlin Distributing C…OHJul 22, 2026
Class IIOngoingZ-2743-2026

Daavlin Distributing Company recalls Photera 600

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

  • Dermatological Ultraviolet Light
  • Software Design Change
Daavlin Distributing C…OHJul 22, 2026