Olympus Corporation Of The Americas recalls Olympus Glass Body for 194 Ellik Evacuator. Model Number: 190A.
Reason for recall
Some sterilization parameters included in the IFU may not consistently render the devices adequately sterile and therefore should no longer be used to sterilize the devices.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Olympus Glass Body for 194 Ellik Evacuator. Model Number:190AUDI-DI 00821925003538.Model Number
What the firm is doing
Olympus notified consignees on about 06/05/2026 via Courier, 2-day delivery. Consignees were requested to examine inventory for affected units. If affected units are on hand, replace any existing sterilization instructions for these products with those provided with the notification and add these instructions to any existing copies of the IFU. Consignees were also instructed to acknowledge receipt of the customer letter through the web portal. If affected units were further distributed, Olympus instructed consignees to identify those customers and forward the notification to them. An updated notice was sent to consignees on 06/30/2026 with further clarification regarding instructions to be used to sterilize affected devices going forward.
DistributionShow detailsHide
Worldwide distribution - US Nationwide and the countries of Canada, Japan, EU, and Latin America.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2721-2026
- FDA device classification · KQTOfficial FDA classification for this device type
- CFR regulation · 21 CFR 876.4370The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Olympus Corporation of the AmericasSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
GE Medical Systems Information Technologies Inc recalls FRU (Field replaceable unit)
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design
GE Medical Systems Information Technologies Inc recalls PORTRAIT CORE SERVICES SOFTWARE V1.0.5
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design
GE Medical Systems Information Technologies Inc recalls PORTRAIT COMPUTE AND CORE SERVICES V1.2
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design