Orthofix U.S. LLC recalls Orthofix MINIRAIL
Reason for recall
Unilateral external fixation devices may be incorrectly labeled. Minirail M101 was labeled as M103, Minirail M103 was labeled as M101, and Minirail M122 was labeled as M111
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Orthofix MINIRAIL, STANDARD LENGTHENER Model/Catalog Number: M101, Unilateral orthopedic/trauma external fixation deviceGTIN 18032568866636
What the firm is doing
On June 8, 2026, Orthofix issued a Urgent Medical Device Recall Notification, via E.Mail. Orthofix asked consignees to take the following actions: 1.Discontinue use of the affected product lots immediately. 2.Immediately examine your inventory and quarantine any affected product and return them to Orthofix. 3. Complete the attached response form and return it to Orthofix. 4. This notice should be shared with all relevant personnel within your organization and forwarded to any other organization or facility where potentially affected devices may have been transferred.
DistributionShow detailsHide
US: AZ CA CO CT FL GA ID IL IN LA MD MI MO MS MT NC NY OH OR SC SD TX UT VA WA WI OUS: None
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2729-2026
- FDA 510(k) clearance · K113770The device's official FDA premarket clearance record
- FDA device classification · KTTOfficial FDA classification for this device type
- CFR regulation · 21 CFR 888.3030The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Orthofix U.S. LLCSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Tosoh Bioscience Inc recalls ST AIA-Pack Testosterone
The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone resul…
- Testosterones And Dihydrotestosterone Radioimmunoassay
- Component design/selection
Daavlin Distributing Company recalls Photera 600 3-D
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Dermatological Ultraviolet Light
- Software Design Change
Daavlin Distributing Company recalls Photera 600
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
- Dermatological Ultraviolet Light
- Software Design Change