Class IIOngoingZ-2741-2026

Maquet Cardiovascular, LLC recalls The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000

Maquet Cardiovascular, LLCWayne, NJ, United StatesReported Jul 22, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Getinge identified that Vasoview Hemopro 2 devices from Lot 3000547797 were exposed to an out-of-specification sterilization dose during a gamma sterilization run (Run ID 1183-28782A) performed by STERIS at its Whippany, NJ facility between March 31 and April 1 of 2026.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

The Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile, single-use Class II device that was cleared for marketing by FDA under 510(k) number K101274.

Lot / code information

UDI
00607567700406, Serial/
Lot #
3000547797

What the firm is doing

Getinge issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 6/18/2026 via FedEx. The notice explained the issue, risk to health, and requested the following actions be taken: " Immediately identify and quarantine all affected Vasoview Hemopro 2 devices from lot 3000547797. " Do not use affected devices. " If an affected device was used, no further intervention is required. Standard post-operative wound care per facility protocol should be followed. " If any affected devices were transferred within your organization or to another organization, ensure this notice is shared with the appropriate individuals. " Contact Getinge Customer Service 1-888-880-2874 or your Getinge representative to coordinate return of the affected units and replacement product. " Complete and return the enclosed response form to Getinge by e-mailing a copy to recallresponses.qrc@getinge.com.

DistributionShow details

US Nationwide distribution in the states of NY, CA, SC, FL, TX, GA, OR.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-2746-2026

Tosoh Bioscience Inc recalls ST AIA-Pack Testosterone

The Test cup lot F level 1 low-quality control results values for the device were lower than expected on the Automated Immunoassay Analyzer (AIA) which could cause falsely elevated or falsely low total testosterone resul…

  • Testosterones And Dihydrotestosterone Radioimmunoassay
  • Component design/selection
Tosoh Bioscience IncOHJul 22, 2026
Class IIOngoingZ-2744-2026

Daavlin Distributing Company recalls Photera 600 3-D

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

  • Dermatological Ultraviolet Light
  • Software Design Change
Daavlin Distributing C…OHJul 22, 2026
Class IIOngoingZ-2743-2026

Daavlin Distributing Company recalls Photera 600

Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).

  • Dermatological Ultraviolet Light
  • Software Design Change
Daavlin Distributing C…OHJul 22, 2026