Class IIOngoingZ-2767-2026

PIE Medical Imaging B.V. recalls 3mensio Workstation (Vascular Fenestrated) software

PIE Medical Imaging B.V.Maastricht, NetherlandsReported Jul 29, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

3mensio Workstation (Vascular Fenestrated) software

Lot / code information

Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5

What the firm is doing

On 5/7/2026, correction notices were emailed to customers who were asked to do the following: 1) Do not perform measurements on Philips 3D echo datasets containing the doppler information using 3mensio Workstation 10.7, 10.7 SP1 SP5. 2) Ensure that this notice is distributed to all relevant personnel within your organization and to any organization where the affected product may have been transferred to or is in use. 3) PMI recommends follow-up to verify if results of ordered devices are correctly analyzed/calculated. 4) Complete and return the enclosed Customer Acknowledgment Form and return via email to vigilance@pie.nl This issue has been resolved in 3mensio Workstation 10.7 SP6. PMI will support users in upgrading to this version.

DistributionShow details

Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-2794-2026

Karl Storz Endoscopy recalls Flexible Hystero-Fiberscope Kits REF: 11264BBU1

Flexible hysteroscopes indicated for use during diagnostic and operative hysteroscopic procedures may contain an expired single use endoscopic seal for gynecology use, which could impact the assurance of sterility.

  • Hysteroscope (And Accessories)
  • Packaging change control
Karl Storz EndoscopyCAJul 29, 2026
Class IIOngoingZ-2793-2026

Alphatec Spine, Inc. recalls SafeOp Patient Interface

Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication betwee…

  • Evoked Response Electrical Stimulator
  • Software Design Change
Alphatec Spine, Inc.CAJul 29, 2026