PIE Medical Imaging B.V. recalls 3mensio Workstation (Vascular Fenestrated) software
Reason for recall
User measurement value reported by diagnostic bioimaging software intended to measure/visualize cardiovascular structures may be larger (0.8% to 34.4%) than the actual anatomical dimension, if all the following occur: affected software is used, and the user has access only to 3D Echo OR Mitral 3D Echo, and the dataset being analyzed is a Philips 3D echo dataset containing Color Doppler information
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
3mensio Workstation (Vascular Fenestrated) software
Lot / code information
Software Version 10.7. UDIs: (01)08056304455925(11)250423(8012)10.7, (01)08056304455925(11)250523(8012)10.7 SP1, (01)08056304455925(11)250714(8012)10.7 SP2, (01)08056304455925(11)250729(8012)10.7 SP3, (01)08056304455925(11)251016(8012)10.7 SP4, (01)08056304455925(11)251103(8012)10.7 SP5
What the firm is doing
On 5/7/2026, correction notices were emailed to customers who were asked to do the following: 1) Do not perform measurements on Philips 3D echo datasets containing the doppler information using 3mensio Workstation 10.7, 10.7 SP1 SP5. 2) Ensure that this notice is distributed to all relevant personnel within your organization and to any organization where the affected product may have been transferred to or is in use. 3) PMI recommends follow-up to verify if results of ordered devices are correctly analyzed/calculated. 4) Complete and return the enclosed Customer Acknowledgment Form and return via email to vigilance@pie.nl This issue has been resolved in 3mensio Workstation 10.7 SP6. PMI will support users in upgrading to this version.
DistributionShow detailsHide
Worldwide - US Nationwide distribution in the states of CA, NY, MN, MI, TX and the countries of United Kingdom, Australia, France, Italy, Hong Kong SAR, Thailand, Taiwan, Netherlands, Ireland, Spain, Greece, Japan, Belgium.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2767-2026
- FDA 510(k) clearance · K250330The device's official FDA premarket clearance record
- FDA device classification · QIHOfficial FDA classification for this device type
- CFR regulation · 21 CFR 892.2050The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find PIE Medical Imaging B.V.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Karl Storz Endoscopy recalls Flexible Hystero-Fiberscope Kits REF: 11264BBU1
Flexible hysteroscopes indicated for use during diagnostic and operative hysteroscopic procedures may contain an expired single use endoscopic seal for gynecology use, which could impact the assurance of sterility.
- Hysteroscope (And Accessories)
- Packaging change control
Alphatec Spine, Inc. recalls SafeOp Patient Interface
Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication betwee…
- Evoked Response Electrical Stimulator
- Software Design Change
Medline Industries, LP recalls Medline medical convenience kits labeled as: 1) TOTAL JOINT PACK Medline Kit SKU DYN…
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
- Orthopedic Tray
- Process change control