Medtronic Neurosurgery recalls Ventriculostomy Kit
Reason for recall
Ventriculostomy devices, used during ventriculostomy procedures to gain access to the ventricular system, may exhibit endotoxin levels that exceed the acceptable limit for devices that come into contact with cerebrospinal fluid; endotoxin exposure can cause an acute inflammatory process.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Ventriculostomy Kit, REF:46154UDI-DI 00673978947396Affected lot2025-08779
What the firm is doing
On 6/17/2026, recall notices were mailed and emailed to customers asking them to do the following: 1) Identify, quarantine, and return affected devices to the firm. 2) Complete and return the Customer Confirmation Form via email to neuro.quality@medtronic.com 3) This notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. For additional information, email the firm at rs.mnsprodexperiencehelp@medtronic.com
DistributionShow detailsHide
Worldwide - US Nationwide distribution in the states of UT, AL, CA, NC, GA, AR, LA, KY, LA, KY, IL, OK and the country of Costa Rica.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2768-2026
- FDA 510(k) clearance · K915546The device's official FDA premarket clearance record
- FDA device classification · HBGOfficial FDA classification for this device type
- CFR regulation · 21 CFR 882.4300The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Medtronic NeurosurgerySearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Karl Storz Endoscopy recalls Flexible Hystero-Fiberscope Kits REF: 11264BBU1
Flexible hysteroscopes indicated for use during diagnostic and operative hysteroscopic procedures may contain an expired single use endoscopic seal for gynecology use, which could impact the assurance of sterility.
- Hysteroscope (And Accessories)
- Packaging change control
Alphatec Spine, Inc. recalls SafeOp Patient Interface
Intraoperative neuromonitoring device, due to a software issue, may experience loss or interruption of triggered electromyography (tEMG) functionality during use, displaying as system faults, loss of communication betwee…
- Evoked Response Electrical Stimulator
- Software Design Change
Medline Industries, LP recalls Medline medical convenience kits labeled as: 1) TOTAL JOINT PACK Medline Kit SKU DYN…
Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.
- Orthopedic Tray
- Process change control