Medtronic Perfusion Systems recalls EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR
Reason for recall
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
EOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 77522
Lot / code information
- UDI
- 00199150023288
- Lot #
- 0233286436
- UDI
- 20199150023282
- Lot #
0232709428 0232744924 0232745061 0232745092 0232798556 0232862763 0232862778 0232862794 0232862798 0232903216 0232903291 0232903383 0232990983 0233286318 0233286414 0233286436 0233427120 0233528730 02…Show all
0232709428 0232744924 0232745061 0232745092 0232798556 0232862763 0232862778 0232862794 0232862798 0232903216 0232903291 0232903383 0232990983 0233286318 0233286414 0233286436 0233427120 0233528730 0233553623 0233556386 0233585113 0233585114 0232670802
What the firm is doing
An URGENT: MEDICAL DEVICE RECALL notification letter dated June 2026 was sent to customers beginning 6/24/26. Customer Actions: Medtronic requests that you take the following actions: " Review your inventory for listed products using Attachment A. " Immediately identify and quarantine all unused, listed product in your inventory. " Return unused, listed product in your inventory to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return for credit or replacement product. Your Medtronic sales representative can assist you in the return of affected product as necessary. " Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have any affected product in your possession. " Please share this notification with others in your organization as appropriate. If the product listed above has been forwarded to another facility, please notify the facility of this Medtronic Urgent Medical Device Recall. " Please maintain a copy of this communication in your records. If you have any questions regarding this communication, please contact 1-800-854-3570, Option 1, then Option 4.
DistributionShow detailsHide
Worldwide - US Nationwide distribution.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2807-2026
- FDA 510(k) clearance · K031037The device's official FDA premarket clearance record
- FDA device classification · DWFOfficial FDA classification for this device type
- CFR regulation · 21 CFR 870.4210The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Medtronic Perfusion SystemsSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Stryker GmbH recalls Stryker Hoffmann II Connecting Rod Micro 3 x 150mm. Catalog Number: 50796150.
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
- Multiple Component Nail/Blade/Plate Combination Fixation Appliance
- Device Design
Stryker GmbH recalls Stryker Hoffmann II Connecting Rod Micro 3 x 120mm. Catalog Number: 50796120.
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
- Multiple Component Nail/Blade/Plate Combination Fixation Appliance
- Device Design
Stryker GmbH recalls Stryker Hoffmann II Connecting Rod Micro 3 x 90mm. Catalog Number: 50796090.
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
- Multiple Component Nail/Blade/Plate Combination Fixation Appliance
- Device Design