Medtronic Perfusion Systems recalls EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR

Medtronic Perfusion SystemsBrooklyn Park, MN, United StatesReported Aug 5, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • EOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number77622
    UDI-DI 20763000135670
    20 affected lots
    02326015010232601510023260154902326015500232602178023260229002326023020232910952
    +12 more023291095302329109540232910955023324407502332443010233244531023328636202332863740233286390023340903402334160710233528731

What the firm is doing

An URGENT: MEDICAL DEVICE RECALL notification letter dated June 2026 was sent to customers beginning 6/24/26. Customer Actions: Medtronic requests that you take the following actions: " Review your inventory for listed products using Attachment A. " Immediately identify and quarantine all unused, listed product in your inventory. " Return unused, listed product in your inventory to Medtronic by contacting Customer Service at 1-800-854-3570, Option 1 then Option 4, and referencing this communication to initiate return for credit or replacement product. Your Medtronic sales representative can assist you in the return of affected product as necessary. " Complete the enclosed Customer Confirmation Form and email to RS.CFQFCA@medtronic.com. This form must be returned even if you do not have any affected product in your possession. " Please share this notification with others in your organization as appropriate. If the product listed above has been forwarded to another facility, please notify the facility of this Medtronic Urgent Medical Device Recall. " Please maintain a copy of this communication in your records. If you have any questions regarding this communication, please contact 1-800-854-3570, Option 1, then Option 4.

DistributionShow details

Worldwide - US Nationwide distribution.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.