Instrumentation Laboratory recalls Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/C…
Reason for recall
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000. portable blood analyzerUDI 01952260009246 affected lots26012062260120652601206926012070202600024015859
What the firm is doing
Werfen dba Instrumentation laboratory notified consignees on about 05/08/2026 via letter and/or email. Consignees were instructed to discontinue use of the affected BCR wands immediately, manually enter patient IDs when performing testing on the GEM Premier 5000 and/or 7000 until the BCR wands have been replaced, contact Werfen Service at 1-800-955-9525 to arrange for prompt replacement of the affected BCR wands, post this notification on or near affected GEM Premier 5000 and/or 7000 instruments, and inform all laboratory staff of this notification. Consignees were also instructed to forward the notification to all affected locations within each facility, retain a copy of the notification for records, and complete and return the Mandatory Response Tracking Form within 10 days of receipt. In addition to the tracked customer notification, a Service Information (SI) 2026-09 was issued on May 11, 2026, for service personnel to track the destruction of the affected BCR wands. The Service Information includes a destruction form for service personnel to document the destruction of the affected BCR wands and to return the form to Regulatory for tracking purposes.
DistributionShow detailsHide
US Nationwide distribution.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2822-2026
- FDA 510(k) clearance · K223608The device's official FDA premarket clearance record
- FDA device classification · CHLOfficial FDA classification for this device type
- CFR regulation · 21 CFR 862.1120The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Instrumentation LaboratorySearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Stryker GmbH recalls Stryker Hoffmann II Connecting Rod Micro 3 x 150mm. Catalog Number: 50796150.
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
- Multiple Component Nail/Blade/Plate Combination Fixation Appliance
- Device Design
Stryker GmbH recalls Stryker Hoffmann II Connecting Rod Micro 3 x 120mm. Catalog Number: 50796120.
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
- Multiple Component Nail/Blade/Plate Combination Fixation Appliance
- Device Design
Stryker GmbH recalls Stryker Hoffmann II Connecting Rod Micro 3 x 90mm. Catalog Number: 50796090.
Potential cracking of the Hoffmann II Carbon Connecting Rods following cleaning and sterilization.
- Multiple Component Nail/Blade/Plate Combination Fixation Appliance
- Device Design