Nuclein LLC recalls Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/C…
Reason for recall
Due to a software issue which may result in false co-infection (e.g., Positive for Flu A, when SARS-CoV-2 is present).
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH¿ Rapid PCR Instrument (DASH Instrument) is an automated, fluorescence-based, reverse-transcription polymerase chain reaction (RT-PCR) instrument that extracts specimens and performs nucleic acid amplified tests (NAAT) in DASH test cartridges for the qualitative detection of infectious diseases. The DASH Instrument is intended for use in conjunction with DASH Rapid PCR System tests to aid in the diagnosis of infectious diseases. Component: No.
Lot / code information
- Lot #
- Code: SN: C01250010
- UDI
- (01)00850063063005(11)250218(21)C01250010; SW: 6.0.2 SN: C02250006
- UDI
- (01)00850063063005(11)250225(21)C02250006; SW: 6.0.2
What the firm is doing
On 06/25/2025, a "MEDICAL DEVICE RECALL" Letter was emailed to customers informing them that, Nuclein identified an incorrect software file on a small number of DASH instruments that creates a possibility for the affected instruments to make incorrect calls that indicate coinfection in the patient tested. On 07/16/2026, the firm emailed an updated "URGENT: MEDICAL DEVICE RECALL" Letter. -Customer were informed regarding updates to the reason for the recall and potential health risks. -Customer were instructed to: All affected DASH Instrument identified in Section 6 (below) have already been replaced. Please complete the Acknowledgement and Receipt Form below and either email the completed form to customersupport@nuclein.com or mail to the address provided at the bottom of the Acknowledgement and Receipt Form. For Questions - contact Nuclein at Phone: (888) 992-DASH on Monday through Friday, 8:30 AM to 5:30 PM, CST. Email: customersupport@nuclein.com Website: https://www.Nuclein.com/contact-us/
DistributionShow detailsHide
US Nationwide distribution in the state of Texas.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2861-2026
- FDA 510(k) clearance · K241652The device's official FDA premarket clearance record
- FDA device classification · NSUOfficial FDA classification for this device type
- CFR regulation · 21 CFR 862.2570The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Nuclein LLCSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Contego Medical, Inc recalls Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Em…
Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction;…
- Carotid Stent
- Process control
Freudenberg Medical, LLC recalls Blom-Singer Classic Indwelling Voice Prosthesis
Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.
- Laryngeal (Taub) Prosthesis
- Labeling mix-ups
Freudenberg Medical, LLC recalls Blom-Singer Classic Indwelling Voice Prosthesis
Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.
- Laryngeal (Taub) Prosthesis
- Labeling mix-ups