Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-2897-2026

United Orthopedic Corp. recalls Brand Name: U2 Total Knee System Product Name: Femoral offset bushing

United Orthopedic Corp.Kaohsiung, TaiwanReported Aug 12, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Due to a design issue causing the incorrect engraved numbering on implant that could result in the incorrect orientation of the implant.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Brand Name: U2 Total Knee System Product Name: Femoral offset bushing, 0mm and 4mm Model/Catalog Number: 9323-5310 Software Version: N/A Product Description: Manual Orthopedic Surgical Instrument Component: No, N/A

Lot / code information

Lot #
Code: Product Number 9323-5310
UDI
04719886908060
Lot #
25FC58

What the firm is doing

On 03/09/2026, the firm distributed an "URGENT: MEDICAL DEVICE RECALL" Letter informing customer that due the fact that the Femoral offset bushing in the affected devices is engraved with the incorrect numbering, and the rotational direction does not align with the intended orientation. Customers are instructed to: 1. Examine inventory immediately to locate the affected product whose quantities and lot numbers are listed in the letter. Note the product may have already been returned. 2. Cease all use of these devices immediately and quarantine all affected devices in their territory for immediate return. 3. Customer Service will be contacting customers regarding replacements. If customers still have the product, they are requested to immediately return the affected product using the prepaid airbills on RMA XXXX . a. Write the RMA number on the outside of the box clearly and legibly. b. Ship recalled product via Fed Ex standard overnight to: UOC USA INC. Attention: RMA# XXXX 16601 Hale Ave. Irvine, CA 92606 (Attached Shipping Label) 4. Additionally, if customers has further distributed the recall product, please provide that customer with a copy of this recall notification and assist them with the return of the product. 5. Complete and return the following page according to the instructions. For questions - contact UOC Sr. Director of Sales Operations at 949.328.3366.

DistributionShow details

US Nationwide distribution in the states of Texas and Michigan.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-2910-2026

Freudenberg Medical, LLC recalls Blom-Singer Classic Indwelling Voice Prosthesis

Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.

  • Laryngeal (Taub) Prosthesis
  • Labeling mix-ups
Freudenberg Medical, L…CAAug 12, 2026
Class IIOngoingZ-2909-2026

Freudenberg Medical, LLC recalls Blom-Singer Classic Indwelling Voice Prosthesis

Incorrect labeling- The labeled box for the Large Esophageal Flange contains a Large Esophageal and Tracheal Flange, and the labeled box for the Large Esophageal and Tracheal Flange contains a Large Esophageal Flange.

  • Laryngeal (Taub) Prosthesis
  • Labeling mix-ups