Class IIOngoingZ-2936-2026

BioFire Diagnostics, LLC recalls BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based…

BioFire Diagnostics, LLCSalt Lake City, UT, United StatesReported Aug 19, 2026 · 7 days ago
Legal News Analyst ·

Reason for recall

Due to increase in false negative results

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based multiplexed nucleic acid test, in vitro diagnostic
    UDI-DI code
    2 affected lots
    3VLB251872125

What the firm is doing

On 06/24/2026, an "URGENT MEDICAL DEVICE CORRECTION" Letter was emailed to customers informing them that bioMrieux received complaints of false-negative results associated with BIOFIRE RP2.1 kit lot 1872125 / pouch lot 3VLB25. An internal investigation confirmed the presence of incorrect PCR1 primers in a subset of pouches within BIOFIRE RP2.1 kit lot 1872125 introduced during reagent manufacturing. This field safety notice was issued because the incorrect PCR1 primers will lead to pouch pathogen assay negative results with passing control assays. Customers are instructed to: "Examine your inventory for the lot identified in this field safety notice. "Cease use and discard any remaining product from this lot in your possession. "Confirm the quantity of kits that have been destroyed, previously used, or returned on the attached Acknowledgement of Receipt Form. "If you have further distributed this product, please identify any recipients and notify them at once. "Complete the accompanying Acknowledgement of Receipt Form and return to bioMrieux so that bioMrieux may acknowledge your receipt of this notification. "bioMerieux will be providing pouch credit for affected products. For Questions - contact local Customer Service Representative at (800) 682-2666. For non-urgent inquiries, contact the Technical Support Center by email at biofiresupport@biomerieux.com.

DistributionShow details

U.S. Nationwide distribution in the states of AL, AR, IA, IL, KS, LA, MN, MO, TX, and WI.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.