BioFire Diagnostics, LLC recalls BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based…
- Multi-Target Respiratory Specimen Nucleic Acid Test Including Sars-Cov-2 And Other Microbial Agents
- Under Investigation by firm
Reason for recall
Due to increase in false negative results
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- BIOFIRE Respiratory Panel 2.1 (RP2.1) REF 423742 UDI code: 00815381020529 PCR-based multiplexed nucleic acid test, in vitro diagnosticUDI-DI code2 affected lots3VLB251872125
What the firm is doing
On 06/24/2026, an "URGENT MEDICAL DEVICE CORRECTION" Letter was emailed to customers informing them that bioMrieux received complaints of false-negative results associated with BIOFIRE RP2.1 kit lot 1872125 / pouch lot 3VLB25. An internal investigation confirmed the presence of incorrect PCR1 primers in a subset of pouches within BIOFIRE RP2.1 kit lot 1872125 introduced during reagent manufacturing. This field safety notice was issued because the incorrect PCR1 primers will lead to pouch pathogen assay negative results with passing control assays. Customers are instructed to: "Examine your inventory for the lot identified in this field safety notice. "Cease use and discard any remaining product from this lot in your possession. "Confirm the quantity of kits that have been destroyed, previously used, or returned on the attached Acknowledgement of Receipt Form. "If you have further distributed this product, please identify any recipients and notify them at once. "Complete the accompanying Acknowledgement of Receipt Form and return to bioMrieux so that bioMrieux may acknowledge your receipt of this notification. "bioMerieux will be providing pouch credit for affected products. For Questions - contact local Customer Service Representative at (800) 682-2666. For non-urgent inquiries, contact the Technical Support Center by email at biofiresupport@biomerieux.com.
DistributionShow detailsHide
U.S. Nationwide distribution in the states of AL, AR, IA, IL, KS, LA, MN, MO, TX, and WI.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2936-2026
- FDA device classification · QOFOfficial FDA classification for this device type
- CFR regulation · 21 CFR 866.3981The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find BioFire Diagnostics, LLCSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) BASIC PACK
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Eye Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) KIT
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Orthopedic Tray
- Nonconforming Material/Component
Cardinal Health 200, LLC recalls Cardinal Health Presource Kits: 1) CHI STD NEURO CRANI OPT 2
Procedure kits contain MEDICOM Ritmed Neurological Sponges that were recalled due to unexpected variability in endotoxin levels.
- Neurological Tray
- Nonconforming Material/Component