Class IIOngoingZ-2941-2026

BD SWITZERLAND SARL recalls 60 In Non-DEHP Minibore Extension Set RES ME2010 UDI-DI code: 10885403236235 MaxPlus…

BD SWITZERLAND SARLEysins, SwitzerlandReported Aug 19, 2026 · 7 days ago
Legal News Analyst ·

Reason for recall

Products may not have received adequate sterilization processing.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • 60 In Non-DEHP Minibore Extension Set RES ME2010 UDI-DI code: 10885403236235 MaxPlus Ext Set w/Needleless and Y-Site REF MPX5305-C UDI-DI code: 10885403274619 9 IN Extension Set w/Max Y REF MPX5300-C UDI-DI code: 10885403237218 Intravascular Administration Set and Extension Set used to administer fluids
    UDI-DI codeREF MPX5305-C
    7 affected lots
    26025260MPX5305-C1088540327461926035455MPX5300-C1088540323721826055528

What the firm is doing

On 06/16/2026, an "URGENT: Medical Device Recall" Letter was emailed to customers informing them that, BD has identified a mechanical issue with internal sterilization equipment resulting in potential underdosing of product during processing, which may have resulted in product not being sterilized correctly. Customer are instructed to: "Immediately discontinue use of the affected lot(s). "Check all inventory locations within your facility and destroy all affected units in accordance with your facility s process. If destruction is not immediately possible at your facility, then product should be quarantined until disposal. "If product is currently in use, replace with non-impacted product as soon as possible. 1. Provide a copy of this notification to any customers and/or facilities you may have distributed devices to. 2. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 3. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 4. Report any complaints experienced with the use of this product to BD via the North American Regional Complaint Center: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT Email: productcomplaints@bd.com For: Product Complaints, Technical Questions Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaints when prompted Mon Fri 8:00am and 5:00pm CT

DistributionShow details

U.S.: AZ, CA, CO, FL, GA, IL, IN, KS, MD, MI, MN, MO, MS, NC, NV, NY, OH, OK, OR, PA, SD, TN, TX, UT, VA, WA O.U.S.: Canada

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.