Siemens Healthcare Diagnostics Products Ltd. recalls Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF(s):L2kSY2/10489085
Reason for recall
Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient results being reported as nonreactive.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Siemens Healthineers Immulite 2000 Systems SYP Syphilis Screen REF(s):L2kSY2/10489085; L2KSY6/10488319 IVD, RxOnly
Lot / code information
- Lot #
- Code: D387/
- UDI
- (s):L2KSY2-00630414999791; L2KSY6 - 00630414999807
What the firm is doing
On July 13, 2026, Siemens Healthineers issued a Urgent Medical Device Correction to affected consignees via FedEx. Siemens asked consignees to take the following actions: 1. Review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2.Discontinue use and discard the affected kit lot. 3. Review your current inventory to determine your laboratory s replacement needs and provide this information to Siemens Healthineers for reporting to the authorities. 4. Complete and return the Field Correction Effectiveness Check Form attached to this letter within (30) days. 5.Retain this letter with your laboratory records and forward this letter to those who may have received the affected product(s).
DistributionShow detailsHide
Worldwide - US Nationwide distribution in the states of FL, NJ, NY and the countries of Austria, Azerbaijan, Belgium, Brazil, Chile, Colombia, Cura-ao, St.¿Eus Cyprus, France, Germany, Italy, Kazakhstan, Latvia, Lithuania, Netherlands, Romania, Russian¿Fed., Saudi¿Arabia, Slovakia, Slovenia, South¿Africa, Spain, U.A.E., Ukraine, United¿Kingdom, Uzbekistan.¿¿¿¿¿
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2966-2026
- FDA 510(k) clearance · K091361The device's official FDA premarket clearance record
- FDA device classification · LIPOfficial FDA classification for this device type
- CFR regulation · 21 CFR 866.3830The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Siemens Healthcare Diagnostics Products Ltd.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
GE Medical Systems Information Technologies Inc recalls FRU (Field replaceable unit)
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design
GE Medical Systems Information Technologies Inc recalls PORTRAIT CORE SERVICES SOFTWARE V1.0.5
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design
GE Medical Systems Information Technologies Inc recalls PORTRAIT COMPUTE AND CORE SERVICES V1.2
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design