Beckman Coulter Inc. recalls Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower:…
Reason for recall
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers
Lot / code information
- Part / Item #
- C63522, UDI/DI 14987666545089; 2)
- Part / Item #
- C63521, UDI/DI 14987666545072; 3)
- Part / Item #
- C13252, UDI/DI 15099590742331
What the firm is doing
Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice (dated 7/10/26) to its consignees on 7/15/2026 via USPS and/or email. The notice explained the issue, impact, and requested the following: The customers are advised to: " Do not change any of the 25 predefined calculated rules. In the case that the default expression does not meet the laboratory s requirements, Beckman Coulter has developed a Job Aid, which outlines the workflow required if laboratories require calculated test rules outside of those predefined calculated rules. " Consult with the Laboratory Manager and/or Medical Director to determine whether any of the predefined rules in the table were modified from the default expression during or following analyzer installation. o If no default expression was modified, there is no action to be taken. o If a predefined calculated rule was modified as compared with the default expressions in the table, follow the steps in the letter to configure this rule as a custom calculated rule in System configuration and to disable the predefined calculated rule that was modified. " Share the letter with the Medical Director to determine if performing a retrospective review of results is necessary. " Post the letter on or near the affected analyzers. For questions regarding this notice, please contact our Customer Support Center. " From our website: http://www.beckmancoulter.com " By phone: call 1-800-854-3633 in the United States and Canada. o Outside the United States and Canada, contact your local Beckman Coulter representative.
DistributionShow detailsHide
Worldwide distribution - US Nationwide.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2973-2026
- FDA 510(k) clearance · K242190The device's official FDA premarket clearance record
- FDA device classification · JJEOfficial FDA classification for this device type
- CFR regulation · 21 CFR 862.2160The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Beckman Coulter Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
GE Medical Systems Information Technologies Inc recalls FRU (Field replaceable unit)
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design
GE Medical Systems Information Technologies Inc recalls PORTRAIT CORE SERVICES SOFTWARE V1.0.5
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design
GE Medical Systems Information Technologies Inc recalls PORTRAIT COMPUTE AND CORE SERVICES V1.2
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design