Medtronic MiniMed, Inc. recalls MiniMed 670G Insulin Pump
Reason for recall
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- MiniMed 670G Insulin Pump, Model: MMT-1781MMT-1782UDI-DI Serial7 affected lotsMMT-1781/00763000404536/NG2665223HNG2748200HNG2748664HNG2755237HNG2716176HNG3547582HMMT-1782/00763000404543/NG2768190H
What the firm is doing
On 7/13/2026, field safety notices were sent to customers asking them to do the following: 1) Stop using the loaner pump and contact the firm to arrange for loaner pump return. 2) Contact your Healthcare Professional (HCP) to ensure you have access to backup insulin therapy supplies or alternative plans, which may include short- and long-acting insulins, syringes, with dosage instructions from your HCP. 3) Complete and return the customer acknowledgement form. Questions can be directed to the firm's Tech Support: 1-800-646-4633
DistributionShow detailsHide
International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-2981-2026
- FDA device classification · OZPOfficial FDA classification for this device type
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Medtronic MiniMed, Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
GE Medical Systems Information Technologies Inc recalls FRU (Field replaceable unit)
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design
GE Medical Systems Information Technologies Inc recalls PORTRAIT CORE SERVICES SOFTWARE V1.0.5
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design
GE Medical Systems Information Technologies Inc recalls PORTRAIT COMPUTE AND CORE SERVICES V1.2
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…
- Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
- Device Design