Class IIOngoingZ-2991-2026

Medline Industries, LP recalls Medline Hudson RCI

Medline Industries, LPNorthfield, IL, United StatesReported Sep 2, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Medline Hudson RCI, DBD-CIRCUIT,VENT,INSPIRATORY,HEATED,NEO, Medline Product Number/SKU HUD78019 (Teleflex Product Number/SKU 780-19); breathing circuit

Lot / code information

Lot #
codes

What the firm is doing

Medline issued an Urgent Medical Device notice it its consignees on 7/13/2026 via email and USPS. The notice explained the issue, potential risk, and provided actions and precautions for healthcare providers. If evidence of connector overheating or thermal damage is observed: " Assess the patient and ensure continued respiratory support. " Replace the affected circuit as soon as clinically appropriate. " Remove the affected circuit from service and retain it for investigation." For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com. Follow up notice sent on 7/27/2026 to clarify the item numbers. *** Updated customer communication was sent out on 8/5/26. Consignees were instructed to destroy affected inventory. The firm has begun distributing corrected product.

DistributionShow details

Worldwide distribution - US Nationwide and the country of Panama.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-3039-2026

GE Medical Systems Information Technologies Inc recalls FRU (Field replaceable unit)

Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings prev…

  • Patient (Without Arrhythmia Detection Or Alarms) Physiological Monitor
  • Device Design
GE Medical Systems Inf…WISep 2, 2026