Baxter Healthcare Corporation recalls Baxter Novum IQ Syringe Pump

Baxter Healthcare CorporationDeerfield, IL, United StatesReported Sep 9, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Baxter Novum IQ Syringe Pump, Loaner, Model Number40800BAXUSL
    UDI-DI 00197886002300

What the firm is doing

On August 11, 2026, firm began sending customers Urgent Medical Device Correction delivered via email and USPS first class mail. Customers were advised to use the same compatible syringe model (brand/size) for all syringes required within a Multi-Syringe VOT workflow. Customers were provided with additional instructions for situations where the syringe does not support the intended flow rate and the user receives a prompt to confirm continuation of the infusion at the maximum or minimum supported flow rate for that syringe after a syringe change.

DistributionShow details

US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.