Medline Industries, LP recalls Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators OPEN HEART…
Reason for recall
Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators OPEN HEART CDS983348AA CDS983348AB
Lot / code information
- UDI
- 10195327385088 Lots 24IBK346 24JBK539; CDS983348AB
- UDI
- 10198459136986 Lots 25CBF097 25HBB232 25HBB761 25HBH209 25HBU907 25JBQ176
What the firm is doing
Firm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
DistributionShow detailsHide
US Nationwide distribution.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-3068-2026
- FDA device classification · OEZOfficial FDA classification for this device type
- CFR regulation · 21 CFR 870.1200The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Medline Industries, LPSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Cardinal Health 200, LLC recalls Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays:…
The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process.…
- Orthopedic Tray
- Process change control
Cardinal Health 200, LLC recalls Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays:…
The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process.…
- Neurological Tray
- Process change control
Cardinal Health 200, LLC recalls Presource Packs Containing CliniClean Pericare Wipes within BD Urinary Catheter Trays:…
The Urinary Catheter Trays contain CliniClean Pericare Wipes which may potentially be contaminated with a lubricant used during the manufacturing process.…
- Hysterectomy Kit
- Process change control