AVID Medical, Inc. recalls Halyard IV Start Kit Non-Sterile REF: KRIV26-02
Reason for recall
Convenience kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which lacks sterility assurance due to to wrinkles in the paper lidding which may breach the seal area.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Halyard IV Start Kit Non-Sterile REF:KRIV26-02UDI 108091604604812 affected lots7/31/202610809160460481
What the firm is doing
On July 14, 2026, Avid Medical Inc. a subsidiary of Owens Minor issued an Urgent Medical Device Recall to affected consignees via email. I Avid Medical Inc. requested consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified. 1a. Discontinue use of the BD ChloraPrep" Clear - 1 mL and FREPP" Clear 1.5 mL Applicators. 1b. Add warning labels to affected AVID kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. 2. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 3. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 4. Complete the attached Recall Response Form and return by July 31, 2026.
DistributionShow detailsHide
US Nationwide distribution in the state of NE.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-3109-2026
- FDA device classification · LROOfficial FDA classification for this device type
- CFR regulation · 21 CFR 878.4370The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find AVID Medical, Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Windstone Medical Packaging, Inc. recalls AMS Convenience Kit: DSAEK Pack - RX
The convenience kits contain .0.9% Sodium Chloride Injection, USP products in Single Dose freeflex bags, which may contain blue Break-off-Parts (BOPs) that have detached from the infusion port during manipulation.
- Disposable Surgical Instrument Kit
- Equipment maintenance
Nipro Renal Soultions USA, Corporation recalls Citric Complete Liquid Citric Acid Concentrate REF: LC+200-D
Hemodialysis concentrates contain high endotoxin levels
- Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
- Process control
Nipro Renal Soultions USA, Corporation recalls Citric Complete Dry Citric Acid Concentrate REF: DCA+200-25 & DCA+203-25
Hemodialysis concentrates contain high endotoxin levels
- Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
- Process control