GE Healthcare recalls Centricity PACS V2
Reason for recall
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Centricity PACS V2, V3, V4, V6, V7, software product is intended for the storage, reading, diagnostic review, analysis, annotation, distribution, printing, editing, and processing of digital images and data acquired from diagnostic imaging devices, All versions
Lot / code information
- UDI
V4 - 00840682124430, V6 - 00840682104807, V7 - 00840682145572; System IDs: CARTI02312611 CARTI22214087T 928757VPASW 928757VPBSW 602344PVSW2 602344REPSITEB MARICOPAC02308029 602344TIMSVM 441236VPSW 805…Show all
V4 - 00840682124430, V6 - 00840682104807, V7 - 00840682145572; System IDs: CARTI02312611 CARTI22214087T 928757VPASW 928757VPBSW 602344PVSW2 602344REPSITEB MARICOPAC02308029 602344TIMSVM 441236VPSW 805395PVSW 805395PVTSW 310349PACSSWPV UCLAMC2259090 310349PACSSWBPV UCLAMC2259063 916781CPACSSRWT PALMDALEM01810792 PALMDALEM16023955 PALMDALEM20508297 909788HCAPACS1 916781CPACSCKS 916781CPACSBCKS GSHCACPSW GSH320037250 GSHCACTPSW RANCHO02255039 RANCHOPISU2 RANCHOPTESTISU 805370HCALR VMPACS69 LOSROBLES19666790 MCANPACSW CENPACSW 303584HCAPVSW 303584REPBSW 303584HCAPVSWTESTIMS 970476PVSW VVMC15111181 VVMC27255473 302674PACSSW 302674PACSBSW 302736SAEATSW JFKSW JFKR19980220B JFKSWOPT 727462PVSW 727462PVBSW MPMH2259089T 772461HCAPACSW 772461HCAPACSW1 772461HCAPACSWT 352333PACSSW NFRH19663978B 904202BMCDSW 904202BMCSSW 904202BMCDTSW MH826898334 407518COREPACS 407518CPREPB KRMCOREPACSSW KRMC19478073 KRMC25216192 615344HCASW PCMC420084442 WFLHIV01 WESTFLORI19661467 772335PACSSW 772335PVSW 850875PVSW2 904819PVSW 904819PV2SW 904819PVTSW 727519PVBSW 850325PACSCORE CRMC820084433 888663CPBSW 727519PVSW 888663CPTSW 888663CPB2SW 727519PVTSW VAMAF02345966 727519PVTSW-MOCK 407518SCPACSHC 850325HCASWOPT WFLHIVT01 PCMC42259036 352333PACSTSW WELLINGTON01810730 WELLINGTON19678279T WELLINGTON20036598 MJPACSW TRMPACSW TRMNC20021605 100115-8-Centricity PACS-21222171 TRMTESTPACS Doctors Ho22222417 MJPACDRSW 706651HCACORE 912466PVSW SGHS2259039 912276PVSW 912276PVTSW 706278PACSSWPV 706278PACSSWBPV 706278PACSSWTPV VSMQL01987331 404325PACSSWPV 406238PVSW 641422PVSW PROSPECT2259084 PROSPECT22128566 319272SCPSWULP 319272CPDRSW 319272CPDRBSW 319272SCPTSWULP 208COREPVSW EID19660790 208CORETIMS NORTHWEST25297966T 773880PVSW 795079PACSEESW1 795079PACSEESW2 312996PVSW2 JSHPACSSW 312996PVSW1 312996PVSWT1 312996PVSWT2 795079PACST JHSJHOCC02259030 773880PVBSW 773880PVTSW JSHPACSTSW 815285PACUV100 ADVPACSTSW1 847437PVSW 847437DCPVSW 847437SAPVTSW LaitekDataMigrationCPACSLabCPACS 847990PVSW S046PACSSWPV S801PACSSWPV S046PACSSWTPV ADVPACSSW ADVPACSBSW 219325PVSW1 219325PVBSW1 219325PVTSW1 765983PVSW 765983PVSW2 RHHCS02253964T 812232PACSPV 812232PACSTPV THRH20084364 FRNKHAPACSPV FRNKHAPACSTPV FRANKFORT19633100 RRMCPACSW RGMCPACSTW RRMC120080697 VTP01810617T VHT02334235P VTP19617532 985230PVSW NOMC2259032 CFO18227240 985230PVSWS 504903CPACSSW 504903CPACSTSW 504903CPACSIMS 781979PVSW 781979PVBSW 413447BCPVSW 413447HCPVSW 413447PVSW 413748PVSW 413748PVTSW 413748PVBSW 301698PVSW 301698PVTSW NORTHE02315652 NORTHE02370307T 517353PACSEESW 517353PACSEETSW 906387VPSW 248858PVASW 248858PVBSW 248858PVTSW 507288PVSW 507288PVSWT-TESTING OMCR19625746 651855PVSW 651855PVBSW 612831PVTSW 660885PVSW 660885PVTSW S011PACSSWPV S011PACSBSW S011PACSTSW S009PACSSWPV S009PACSSWBPV S009PACSSWTPV2 GPMCCOREPACS 601288PACSSWPV DBCSW 406345SCPACSTSW 704660PVSW 704660PVTSW 402484PACSSW AMIM02031832 609652PACSSW 609652PACSBSW 609652PACSTSW VHGPVSW VHGPVBSW VHGPVTSW UMDNJPVPSW UMDNJPVPBSW UMDNJPVPTSW 505609PVSW SJRMIMST20655929 VSMAI01812924 702733CPACSSW SHPACSW SUNPACSW VALLEYHEA14460287 SUNPACSREPB 20910879CPACST NORTHERNN20284710 NORTHERNN20838223 NORTHERNN20838225 STO01440342 BLH01442255TST BLH01451970 STO02157029 STO02157030T2 718206PVSW 718206PVTSW 2435-1-Centricity PACS-20608371 718206REPBSW 315798PVSW1 315798PTISU MSKPVSW NEWYORK01976546 MSKPVBSW MSKPVTSW2 MSKPVTSW3 ERM16295285 718920PVSW 718920PVTSW1 516763PVSW 516763PVBSW 516763PVTSW1 STO18806599 914357PVSW 914357PVBSW 914357PVTSW 315470CPACSASW 315470CPACSBSW 315470CPACSTSW 914681PVSW 914681PVTSW 718920PVTSW2 330543PACSSWPV CHPA2259061 CHPA2259088T 614234PVSW 614722PVSW 614722PVBSW SAH52259083 SAH52259110T 740348PACSSW 740348PACSBSW 740348PACSTSW 419383PVSW MUOO2259017 419383PVSWB STMARYS-TISU STMARYS-PISU S040PACSSWBPV S040PACSSWAPV S040PACSSWTPV 503257PACSSWPV UHS101383446 215913CPBSW 484526PVSW1 484526PVTSW 484526CPREPB 215710PVSW SMMC14898807 QUANTUMIM14640640A QUANTUMIM14776702B QUANTUMIM14776726T 570523PVSW EVANVTPACSIMS 724728SSASW 724728SSABSW 724728SSATSW2 412575CPACSSW 412623CPBSW 215453PACSSWPV GVH102398920T 401444PVBSW 401444PVSW 401444PVTSW AIMC01810707 AIMC19632995 864716PACSSWPV 864716PACSSWT 864716PACSSWBPV 803461PACSSWPV SCOA2259023 843347PVSW 843347PVTSW CNOS02186078 TEST PQM 423493PACSW 423493REPB WTN00104692T WTN2259070 WTN2259097 423837HCASWSU 423431PVSW 865977SSASW 865977SSTSW 806354CPACS 806354CPACSB 806354CPACST7 972906GSPACSTSW 361694VPSW DRISCOLLC02257225TIMS 214820EEPVSW1 SCW2259068 FORTDUNCA01810773 FORTDUNCA20854738T FORTDUNCA21337445 DSMCCOREPACSSW DSMCCOREPACSDRSW DSMC-22886533 972906GSPACSSW HCA-NORT02332691 MEDIC01812756 281274MSLPVSW 713441MHPVSWA 713704MHMSWA 713704MHMSWB 713704MHMSWTA 713704MHMSWTB 713741MHHPVSWB 832522MWHPVSW DHOL01810756 DHOL15291375T DHOL20631410 DHOL23253841 214820EEPVSW2 254724PVSW SCW2259095 361572SSASW CMDC2259015 801DASSW 801DSWSP1 801SOFTWARE 801SOFTWAREOPT SMH820071869 SMH8-26535851 SMH8-26535857 LGMSW LGHI20084372 STSP21736781 802775SSASW 802775SSATSW4X 206292PACSSWPV 206545EESW UWH20021873REPB 414465PVSW 414465PVTASW 414465PVTBSW 414465PVBSW 304487PVSW PCH12259035-TISU DZ1906PP02 DZ27??PP03 DZ2112CC01 HCIT1555339PACS HCIT4825134PACS HCIT4691855PACS HCIT5184728CPACS HCIT4859945PACS 0910553037 0910553037 ANZVICCAUL ANZVICSAND AH0104CS01 AH5045CS01 BH1038PC01 22303605 010120ST03 010120ST52 300013ST11 HCIT4691866CPACS HCIT4691855CPACS HCIT4759526PACS 4863935_CPACS HCIT7102800PACS HCIT4852374PACS-IMS HCIT5245250CPACS HCIT4860936PACS HCIT5317135CPACS HCIT4290686PACS HCIT5245275CPACS HCIT5471924PACS HCIT7104315PACS HCIT5245248CPACS HCIT5245248TESTCPACS HCIT7229698PACS 604520REPSITEA 604520TESTIMS 604586REPSITEB 709737PACS 709752IMSAsw 709752IMSBsw 709752GEPACS 709752DCSDAS02sw 709752DCSDAS04sw 709752TESTIMSsw 709752REPSERVERsw 709752DCSDAS01sw 709752DCSDAS03sw 709752DASNLA01 709752DASNLA02 709752DASNLA03 709752DASNLA04 709752DASNLA05 709752DASNLA06 709752DASNLB01 709752DASNLB02 709752DASNLB04 709752DASNLB05 709752DASNLB06 709752DASNLB07 709752DASN …
What the firm is doing
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your systems by following the instructions below. 1. Do not perform any new inbound data (DICOM) migration projects. Please contact GE HealthCare before starting any new migration projects. 2. If using Enterprise Archive with Core Services Hub/Workflow Core Services, when reviewing or exporting prior imaging studies after migration to Enterprise Archive, verify that the study and images within the study are associated with the correct patient as follows: At the Study level: " Compare name, date of birth, and patient identifiers only if burned-in or embedded in the image. Do not rely on patient information in the image overlays that can be toggled on and off. " If burned-in or embedded patient identifiers are not available, compare with prior studies based on patient anatomy, implants, or unique anatomical features. At the Image level: " Confirm that the images within the study do not belong to different patients based on acquisition date and time or other patient identifiers. Please maintain any previous source systems or migration servers pending correction of this issue by GE HealthCare. Please ensure that all potential users in your facility are made aware of this safety notification and the recommended actions above. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 85492 Digital Response Form or print, fill out manually, scan, and email to recall.85492@gehealthcare.com. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
DistributionShow detailsHide
Worldwide - US Nationwide distribution.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-3168-2026
- FDA 510(k) clearance · K043415The device's official FDA premarket clearance record
- FDA 510(k) clearance · K110875The device's official FDA premarket clearance record
- FDA device classification · LLZOfficial FDA classification for this device type
- CFR regulation · 21 CFR 892.2050The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find GE HealthcareSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
PI CARDIA LTD recalls Pi-Cardia Inc. ShortCut REF: Z-A-20.003-US
Transfemoral catheter, positing arm may detached due to weld issue.
- Percutaneous Catheter For Cutting Or Splitting Heart Valve Leaflets Concomitant To Transcatheter Valve Procedures
- Process control
ICU Medical, Inc. recalls CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-…
Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or a…
- I.V. Fluid Transfer Set
- Process control
F & A Foundation recalls Reign Medical Clench Staple System (Compression)
Bone fixation staple requires a new 510(k); the performance characteristics of this device have not been adequately established.
- Bone Fixation Staple
- No Marketing Application