Howmedica Osteonics Corp. recalls TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manuf…
- Calcium Phosphate Non-Porous Metal / Polymer Uncemented Semi-Constrained Hip Prosthesis
- Process change control
Reason for recall
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- TRIDENT 0¿ POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)620-00UDI-DI 076133273135812 affected lotsMHANP2MHD1N0
What the firm is doing
Firm began notifying customers on July 30, 2026, via FedEx 2-day air. Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm.
DistributionShow detailsHide
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-3204-2026
- FDA 510(k) clearance · K983502The device's official FDA premarket clearance record
- FDA device classification · MEHOfficial FDA classification for this device type
- CFR regulation · 21 CFR 888.3353The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Howmedica Osteonics Corp.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
PI CARDIA LTD recalls Pi-Cardia Inc. ShortCut REF: Z-A-20.003-US
Transfemoral catheter, positing arm may detached due to weld issue.
- Percutaneous Catheter For Cutting Or Splitting Heart Valve Leaflets Concomitant To Transcatheter Valve Procedures
- Process control
ICU Medical, Inc. recalls CADD Yellow Medication Cassette Reservoir with NRFit connector US Product Codes: 21-…
Reservoirs may exhibit medication leakage at the joint between luer connector and tubing. Issue could result in delay or interruption of therapy, under delivery of medication, exposure to infectious or toxic agents, or a…
- I.V. Fluid Transfer Set
- Process control
F & A Foundation recalls Reign Medical Clench Staple System (Compression)
Bone fixation staple requires a new 510(k); the performance characteristics of this device have not been adequately established.
- Bone Fixation Staple
- No Marketing Application