Bard Peripheral Vascular Inc recalls EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONS…
Reason for recall
Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pressure sensor (code E01-0054000E), which prompts the user to restart and re-calibrate; errors may lead to interruption or prolongation of the biopsy procedure, or the need for additional tissue sampling or repeat biopsy.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with softwareUDI-DI 0080174123522199 affected lotsVTLNPX06VTLNPX0FVTLNPX0DVTLNPX08VTLNPX0KVTLNPX0AVTLNPX0CVTLNPX0L
+91 more
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
What the firm is doing
On 8/20/2026, correction notices were mailed to customer who were asked to do the following: 1.If calibration cannot be completed, clinicians should first follow the troubleshooting instructions provided on the screen and in the Instructions for Use (IFU). If the issue persists, the probe should be replaced and the process repeated. If unsuccessful, an alternative biopsy driver or system should be used, as clinically appropriate. 2. If a pressure sensor error is observed, clinicians should follow the troubleshooting instructions provided on the screen and in the Instructions for Use (IFU). If the error occurs and the system cannot be returned to normal operations, discontinue use of the affected system and complete or reschedule using alternative equipment, as clinically appropriate. No additional follow-up is recommended for patients whose procedures were completed without complication. 3. If affected devices function as intended or if the error codes can be cleared, facilities can continue to use them prior to software update. 4. Post the correction notice on or near affected devices until the software has been updated. 5. Ensure devices are available for the update. 6. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Share this notification with all product users within your facility network to ensure they are also aware of this product correction. If the affected products have been distributed further, please ensure that these entities are contacted and informed of the correction. 7. Complete and return the response form via email to BDRC33@bd.com Firm has developed a software solution to address these issues. A firm representative will contact your facility to arrange the update. If you require assistance, contact the firm via email: productcomplaints@bd.com, or phone: 1-844-823-5433
DistributionShow detailsHide
US Nationwide distribution in the states of KS, SC, CA, TX, GA, NC, PA, SC, ID, MI, KY, CT, IL, VA, MA, WI, NY, IN, TN, OH, UT.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-3275-2026
- FDA 510(k) clearance · K252681The device's official FDA premarket clearance record
- FDA device classification · KNWOfficial FDA classification for this device type
- CFR regulation · 21 CFR 876.1075The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Bard Peripheral Vascular IncSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Revised C-Reactive Protein (RCRP)
The potential for falsely elevated results.
- C-Reactive Protein Test System
- Under Investigation by firm
Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the pl…
- Bone Fixation Plate
- Process control
Karl Storz Endoscopy recalls HOPKINS Telescope 30¿
Due to incorrect inside packaging label for eyepiece.
- Nasopharyngoscope (Flexible Or Rigid)
- Employee error