Bard Peripheral Vascular Inc recalls EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONS…

Bard Peripheral Vascular IncTempe, AZ, United StatesReported Oct 7, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pressure sensor (code E01-0054000E), which prompts the user to restart and re-calibrate; errors may lead to interruption or prolongation of the biopsy procedure, or the need for additional tissue sampling or repeat biopsy.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with software
    UDI-DI 00801741235221
    99 affected lots
    VTLNPX06VTLNPX0FVTLNPX0DVTLNPX08VTLNPX0KVTLNPX0AVTLNPX0CVTLNPX0L
    +91 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

What the firm is doing

On 8/20/2026, correction notices were mailed to customer who were asked to do the following: 1.If calibration cannot be completed, clinicians should first follow the troubleshooting instructions provided on the screen and in the Instructions for Use (IFU). If the issue persists, the probe should be replaced and the process repeated. If unsuccessful, an alternative biopsy driver or system should be used, as clinically appropriate. 2. If a pressure sensor error is observed, clinicians should follow the troubleshooting instructions provided on the screen and in the Instructions for Use (IFU). If the error occurs and the system cannot be returned to normal operations, discontinue use of the affected system and complete or reschedule using alternative equipment, as clinically appropriate. No additional follow-up is recommended for patients whose procedures were completed without complication. 3. If affected devices function as intended or if the error codes can be cleared, facilities can continue to use them prior to software update. 4. Post the correction notice on or near affected devices until the software has been updated. 5. Ensure devices are available for the update. 6. This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. Share this notification with all product users within your facility network to ensure they are also aware of this product correction. If the affected products have been distributed further, please ensure that these entities are contacted and informed of the correction. 7. Complete and return the response form via email to BDRC33@bd.com Firm has developed a software solution to address these issues. A firm representative will contact your facility to arrange the update. If you require assistance, contact the firm via email: productcomplaints@bd.com, or phone: 1-844-823-5433

DistributionShow details

US Nationwide distribution in the states of KS, SC, CA, TX, GA, NC, PA, SC, ID, MI, KY, CT, IL, VA, MA, WI, NY, IN, TN, OH, UT.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-3334-2026

Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the pl…

  • Bone Fixation Plate
  • Process control
Wright Medical Technol…TNOct 7, 2026