Class IIOngoingZ-3276-2026

Volcano Corp recalls Philips IntraSight Plus REF: 797423 and 797417 100-120v-220-240v

Volcano CorpSan Diego, CA, United StatesReported Oct 7, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Software system may associate image data from one patient with another patient's record, potentially resulting in incorrect diagnosis or treatment.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Philips IntraSight Plus REF: 797423 and 797417 100-120v-220-240v,50/60hZ, 1000 VAUDI
    REF 797423

What the firm is doing

On August 6, 2026, Philips issued a Urgent Medical Device Correction to affected consignees. Philips ask consignees to take the following actions: 1. Contact your local Philips Service representative in advance of any changes to the worklist provider, Azurion software, interface configuration, or hospital IT environment, as such changes may affect the Study ID field. Your Philips Service representative will help evaluate any impact and/or reverify as applicable. Otherwise, you may continue to use the device in accordance with the intended use. 2. Please circulate this notice to all users of the IntraSight Plus so they are aware of the issue and associated hazard / harm. Philips also encourages all customers to post this letter on or near the system until Philips has updated its software.

DistributionShow details

Worldwide - US Nationwide distribution in the states of MA, MN, TX, AL, GA, CA and the countries of Belgium, Denmark, Germany, Netherlands, Poland, Spain, Sweden, United Kingdom.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.