Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-3314-2026

Philips North America recalls SmartPath to dStream for 1.5T

Philips North AmericaCambridge, MA, United StatesReported Oct 7, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

When exporting diffusion tensor imaging (DTI) scans from a MR system using the Enhanced MR DICOM format, the exports contain incorrect data (diffusion directions). If the Enhanced MR DICOM export is used in conjunction with third-party software for tractography (e.g. for mapping nerve fiber pathways in the brain), the software will not be able to correctly process the DTI data for reconstruction.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

SmartPath to dStream for 1.5T, SmartPath to dStream for 3.0T, SmartPath to dStream for XR and 3.0T, and SmartPath to Ingenia Elition X; Model number (REF): 1. 781260 (1.5T), 2. 782112 (1.5T), 3. 782146 (1.5T), 4. 782165 (1.5T), 5. 782145 (3.0T), 6. 782164 (3.0T), 7. 781270 (XR and 3.0T), 8. 782113 (XR and 3.0T), 9. 782129 (XR and 3.0T), 10. 782118 (Ingenia Elition X), 11. 782144 (Ingenia Elition X), 12. 782163 (Ingenia Elition X); Software Versions: R11, R12 and R13;

Lot / code information

REF
): 781260 (1.5T)
UDI
00884838095076
Serial #
5479, 5490, 8057, 8118, 8260, 8330 — +40 moreShow all
5479, 5490, 8057, 8118, 8260, 8330, 8392, 8819, 8884, 8954, 10044, 10170, 10236, 11062, 13018, 13029, 13047, 18663, 18872, 20033, 20115, 20164, 20319, 20322, 20413, 21011, 21062, 21133, 21142, 21156, 21278, 21577, 21583, 21621, 21775, 21880, 21969, 22157, 22248, 32049, 32245, 32262, 32286, 32322, 32845, 39024; 2
REF
): 782112 (1.5T)
UDI
00884838098886
Serial #
12002, 21205, 21658, 30031, 32378, 32519, 33338; 3
REF
): 782146 (1.5T)
UDI
00884838108691
Serial #
8935, 13013, 21202, 30308, 32570, 32634, 32679, 32683, 32750, 33074, 33479, 33611, 33714, 39040; 4
REF
): 782165 (1.5T)
Show 26 more code fields
UDI
00884838115767
Serial #
21763, 22170, 32402, 32409, 32971, 33379, 33387, 33571, 39075, 72160, 72190; 5
REF
): 782145 (3.0T)
UDI
00884838108684
Serial #
17377, 17385, 17435, 24010, 24062, 34256, 34282, 38059, 38118, 38127, 38170, 38251, 38278, 38294, 38295, 38302, 38336, 73002, 74004, 74007, 74030; 6
REF
): 782164 (3.0T)
UDI
00884838115750
Serial #
34038, 34254, 34268, 38021, 38046, 38061, 38086, 38090, 38223, 38246, 74020, 74033; 7
REF
): 781270 (XR and 3.0T)
UDI
00884838095083
Serial #
17239, 17246, 17264, 17269, 17283, 17299 — +47 moreShow all
17239, 17246, 17264, 17269, 17283, 17299, 17356, 17357, 17369, 17406, 17416, 17425, 17434, 17464, 17468, 17473, 17514, 17544, 17550, 17564, 17565, 17569, 17589, 17590, 23026, 23033, 24006, 24012, 24023, 24066, 24075, 24087, 24090, 24108, 24110, 34107, 34152, 34154, 34233, 35032, 35047, 35054, 35056, 35087, 38001, 38004, 38023, 38071, 38084, 38089, 38120, 38202, 38300; 8
REF
): 782113 (XR and 3.0T)
UDI
00884838098909
Serial #
17346, 17439, 24039, 24077, 24146, 34131, 34212, 35045, 35063, 35071, 38075, 38197, 38234, 38291; 9
REF
): 782129 (XR and 3.0T)
UDI
00884838105805
Serial #
34055, 38195; 10
REF
): 782118 (Ingenia Elition X)
UDI
00884838099746
Serial #
42072, 42098, 42194, 42298, 42301, 42329, 42340, 71019, 71025; 11
REF
): 782144 (Ingenia Elition X)
UDI
00884838108677
Serial #
42062, 42105, 42323, 42346, 42363, 71087, 71504; 12
REF
): 782163 (Ingenia Elition X)
UDI
00884838115743
Serial #
42299, 42354, 42404, 71390

What the firm is doing

On August 28, 2026, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: 1. To prevent the issue from occurring while Philips is developing the software solution, use the Classic DICOM transfer option for exporting DTI scans. 2. You may continue to use your system(s) in accordance with the intended use. 3. Circulate this notice to all users of this device so that they are aware of the issue and associated hazard/harm. 4. Please retain this Urgent Medical Device Correction letter with your system(s) until the software upgrade is installed; ensure the notice is in a place likely to be seen/viewed.

DistributionShow details

Worldwide distribution - Nationwide and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, Equator. Guinea, Finland, France, French Guiana, Georgia, Germany, Ghana, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mauritius, Mexico, Moldova, Monaco, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Pakistan, Palestine, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Tunisia, T¿rkiye, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Vietnam.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-3334-2026

Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the pl…

  • Bone Fixation Plate
  • Process control
Wright Medical Technol…TNOct 7, 2026