Class IIOngoingZ-3319-2026

Percussionaire Corporation recalls Intrapulmonary Percussive Ventilation

Percussionaire CorporationSandpoint, ID, United StatesReported Oct 7, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number: P5-UC, used with In-Line Valve Instruction Manual

Lot / code information

UDI
S00065/00849436000679, 00849436000709; P5-UC/00849436000686 In-Line Valve Instruction Manual: P20174 Rev D

What the firm is doing

On 8/18/2026, correction notices were emailed to customers and distributors who were informed of the following: The following warning will be added to associated Instructions for Use: Warning: Do not use IPV therapy in conjunction with volatile anesthetic delivery systems. The high-frequency percussive flow generated during IPV therapy may alter the controlled delivery of volatile anesthetic agents within the breathing circuit, resulting in unpredictable anesthetic agent concentrations that may result in serious patient harm. Note: Consult the anesthesia healthcare provider before disconnecting from the system if considering delivering IPV directly to the patient's airway. Customers were asked to do the following: 1) Print and post the new warning for applicable IPV devices throughout your facility to ensure adequate distribution of new information to users. 2) Reinforce adherence to the IPV User Guide and IPV Quick Reference Guide instructions, including ensuring communication and training primary users of IPV therapy, specifically those using it in-line with a ventilator, on the instructions and new warning. Distributors were asked to do the following: - Translate (if applicable) and send the Urgent Medical Device Field Safety Notice to affected end customers. -If the Distributor receives the signed acknowledgement letters from affected customers, please forward them and all related communication to FSCA@sentec.com. If there are any additional questions related to the updated instructions or other questions, please contact FSCA@sentec.com

DistributionShow details

Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-3334-2026

Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the pl…

  • Bone Fixation Plate
  • Process control
Wright Medical Technol…TNOct 7, 2026