SPECTRUM MEDICAL SRL recalls Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger Hi…
Reason for recall
Due to blood-to-glycol leak in the heat exchanger.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Brand Name: Quantum PureFlow Product Name: Quantum PureFlow Standard Heat Exchanger High Flow 3/8 Model/Catalog Number: HX55W-S0 Software Version: N/A Product Description: The Quantum PureFlow Standard Heat Exchanger is intended to be used with a compatible Heater/Cooler system to heat/cool blood during routine cardiopulmonary bypass (CPB) procedures up to 6 hours in duration with a maximum flow rate of 8 l/min. Quantum PureFlow Standard Heat Exchanger is a single use device which allows heating/cooling the blood during extracorporeal circulation. The device is sold sterile in a blister pack in a multiple box of 4 units. Device surfaces in contact with blood are treated with a phosphorylcholine-based biocompatible coating. The device has been designed to be powered by heaters coolers that use water or a glycol-based heat transfer fluid (HTF). Component:N/AAffected lot18051160301178
What the firm is doing
On 07/17/2026, an "URGENT MEDICAL DEVICE RECALL" letter was sent via email to customers informing them that, Spectrum Medical S.r.l. has received customer complaints reporting a leak between the blood and glycol compartments of certain adult heat exchangers. Actions to be taken by customers: During the procedure, the user should monitor the heat exchanger for leakage, which would be evident by a change in colour in the glycol compartment. Below are images of the heat exchanger if a blood-to-glycol leak was to occur. When operating as intended, the glycol compartment should remain clear. If the colour of the glycol compartment changes, then a leak is present between the blood and glycol compartments. If leakage is noted between the blood and glycol compartments of the heat exchanger, the heat exchanger should be immediately replaced in the circuit with a new device. Please return the defective heat exchanger to Spectrum Medical. The contaminated heater-cooler device should also be immediately removed from clinical use and returned to Spectrum Medical for decontamination and inspection. Do not use the heater-cooler with another patient. If leakage has occurred, please contact a Spectrum Medical representative to coordinate device return. If leakage into the water compartment occurs when using a third-party heater-cooler, the customer should immediately remove the heater-cooler from clinical use and contact the heater-cooler manufacturer. Spectrum Medical suggests that for customers that use the heat exchanger in combination with a third-party heater-cooler to contact the manufacturer of the heater-cooler to pre-emptively discuss the steps that should be taken if the device is exposed to blood contamination. For questions or further assistance - contact Regulatory Affairs Specialist at cameron.schmitt@spectrummedical.com
DistributionShow detailsHide
US Nationwide distribution in the states of AL, CA, FL, KS MA, MD, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TN, TX, VA, WI.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-3328-2026
- FDA 510(k) clearance · K220110The device's official FDA premarket clearance record
- FDA 510(k) clearance · K223879The device's official FDA premarket clearance record
- FDA device classification · DTROfficial FDA classification for this device type
- CFR regulation · 21 CFR 870.4240The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find SPECTRUM MEDICAL SRLSearch for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Revised C-Reactive Protein (RCRP)
The potential for falsely elevated results.
- C-Reactive Protein Test System
- Under Investigation by firm
Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the pl…
- Bone Fixation Plate
- Process control
Karl Storz Endoscopy recalls HOPKINS Telescope 30¿
Due to incorrect inside packaging label for eyepiece.
- Nasopharyngoscope (Flexible Or Rigid)
- Employee error