Tandem Diabetes Care, Inc. recalls Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application
Reason for recall
A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.
Affected product
Check your device
Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.
Affected products & lots
- Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: NoUDI-DI 003891524706412 affected lots003891524706412.13
What the firm is doing
On 07/16/2026, an "URGENT MEDICAL DEVICE CORRECTION" communication was sent via certified mail to customers informing them that an unexpected behavior was identified in the t:slim iOS mobile application version 2.13 made available through TestFlight. The issue involves users unable to pair a compatible FreeStyle Libre 3 Plus sensor after updating to version 2.13. If this occurs, the application will incorrectly display an "Incompatible Sensor" message and prevent sensor use. Customer are advised to: 1. Discontinue use of iOS t:slim mobile application version 2.13 2. Uninstall version 2.13 from their mobile device. 3. Return to the current production version of the t:slim mobile application (version 2.12.1) available on the Apple App Store. 4. Scan the compatible FreeStyle Libre 3 Plus sensor with version 2.12.1 when starting their next sensor session. For Questions - contact Tandum Customer Support at 1-877-801-6901 or email Techsupport@tandumdiabetes.com (24/7/365).
DistributionShow detailsHide
US Nationwide distribution.
Use of the product may cause temporary or medically reversible adverse health consequences.
Sources & references
- FDA enforcement recordOfficial FDA database entry for Z-3330-2026
- FDA 510(k) clearance · K201214The device's official FDA premarket clearance record
- FDA 510(k) clearance · K203234The device's official FDA premarket clearance record
- FDA 510(k) clearance · K232380The device's official FDA premarket clearance record
- FDA device classification · QFGOfficial FDA classification for this device type
- CFR regulation · 21 CFR 880.5730The federal regulation governing this device
- FDA recall announcementSearch FDA.gov for the official press announcement
- Find Tandem Diabetes Care, Inc.Search for the manufacturer's official notice
- News coverageRecent reporting on this recall and manufacturer
- openFDA data recordThe raw JSON record this page is built from
Source: U.S. FDA openFDA device enforcement reports. This summary is informational and may lag or differ from the official record. Always verify against the FDA before acting.
Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.
More Class II recalls
Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Revised C-Reactive Protein (RCRP)
The potential for falsely elevated results.
- C-Reactive Protein Test System
- Under Investigation by firm
Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw
Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the pl…
- Bone Fixation Plate
- Process control
Karl Storz Endoscopy recalls HOPKINS Telescope 30¿
Due to incorrect inside packaging label for eyepiece.
- Nasopharyngoscope (Flexible Or Rigid)
- Employee error