Skip to content
RecallWatchMedical Device Safety
Class IIOngoingZ-3330-2026

Tandem Diabetes Care, Inc. recalls Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application

Tandem Diabetes Care, Inc.San Diego, CA, United StatesReported Oct 7, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No
    UDI-DI 00389152470641
    2 affected lots
    003891524706412.13

What the firm is doing

On 07/16/2026, an "URGENT MEDICAL DEVICE CORRECTION" communication was sent via certified mail to customers informing them that an unexpected behavior was identified in the t:slim iOS mobile application version 2.13 made available through TestFlight. The issue involves users unable to pair a compatible FreeStyle Libre 3 Plus sensor after updating to version 2.13. If this occurs, the application will incorrectly display an "Incompatible Sensor" message and prevent sensor use. Customer are advised to: 1. Discontinue use of iOS t:slim mobile application version 2.13 2. Uninstall version 2.13 from their mobile device. 3. Return to the current production version of the t:slim mobile application (version 2.12.1) available on the Apple App Store. 4. Scan the compatible FreeStyle Libre 3 Plus sensor with version 2.12.1 when starting their next sensor session. For Questions - contact Tandum Customer Support at 1-877-801-6901 or email Techsupport@tandumdiabetes.com (24/7/365).

DistributionShow details

US Nationwide distribution.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-3334-2026

Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the pl…

  • Bone Fixation Plate
  • Process control
Wright Medical Technol…TNOct 7, 2026