Class IIOngoingZ-3331-2026

Penumbra Inc. recalls Brand Name: Penumbra System (RED 62 Reperfusion Catheter) Product Name: Penumbra Syste…

Penumbra Inc.Alameda, CA, United StatesReported Oct 7, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Due to manufacturing issues reperfusion catheters may fracture.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Brand Name: Penumbra System (RED 62 Reperfusion Catheter) Product Name: Penumbra System (RED 62 Reperfusion Catheter) Model/Catalog Number: THNDRRED62SKIT Software Version: N/A Product Description: Product common name: Percutaneous catheter A complete product description is provided in Attachment 1 (HHE-2026-002 RED62). Component: Penumbra System
    UDI 00815948024359
    2 affected lots
    00815948024359H00007856

What the firm is doing

On 07/24/2026, an "URGENT: VOLUNTARY MEDICAL DEVICE RECALL NOTIFICATION" was sent via email and FedEx Priority Overnight to customers informing them that Penumbra has identified an issue (frature) in the lot H00007856 of the Penumbra System RED 62 Reperfusion Catheter (RED 62). Actions Customers need to take: " Share this Recall Notification with all users of the product within your facility " Review inventory for the specified lot listed above " If you do have inventory of the specified lot listed above, remove affected units from inventory and place in quarantine for return to Penumbra, Inc. Complete and return the Product Identification/Return Notification to notifications@penumbrainc.com within 3 business days. Your Penumbra Sales Representative will work with you directly to arrange return and credit. " If you do not have inventory of the specified lot listed above, complete and return the Product Identification/Return Notification form to notifications@penumbrainc.com within 3 business days. No further action is required. For questions - contact Penumbra Sales Representative or email at notifications@penumbrainc.com.

DistributionShow details

US Nationwide distribution in the states of AL, CA, CO, FL, LA, NC, NY, TX, UT, WI.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-3334-2026

Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the pl…

  • Bone Fixation Plate
  • Process control
Wright Medical Technol…TNOct 7, 2026