Class IIOngoingZ-3332-2026

Beckman Coulter Mishima K.K. recalls Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU48…

Beckman Coulter Mishima K.K.Sunto-Gun, JapanReported Oct 7, 2026 · 6 days ago
Legal News Analyst ·

Reason for recall

Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

Affected product

Check your device

Enter a lot, serial, REF, or UDI number from your device's label to check it against this recall's listed codes.

Affected products & lots

  • Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B96693 / 14987666542972 B12183 / 14987666540671 The AU480 Chemistry Analyzer measure analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material, and other accessories. This system is for in vitro diagnostic use only
    UDI-DI codesUDI number

What the firm is doing

On 08/03/2026, an "URGENT MEDICAL DEVICE RECALL" Letter was sent via postal mail and email informing customers that CHEMISTRY ANALYZER AU480 with Windows 10 IoT Enterprise 2021 LTSC PCsmay output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. Customers are instructed to: 1. Be sure to create a new index once or more per day at an appropriate time. Refer to 5.1 Start the System of the IFU to complete this action. 2. Beckman Coulter recommends posting this letter on or near the affected analyzers. Resolution - Beckman Coulter has identified the root cause and will resolve the issue by applying a Software solution to the impacted systems. BEC service teamwill contact customers to schedule the Software solution application, and a Field Service Engineer (FSE) will perform the on-site work. For questions - contact Customer Support Center at: http://www.beckmancoulter.com

DistributionShow details

Worldwide - U.S. Nationwide distribution in the states of AR, CA, DE, FL, GA, GU, HI, IA, IL, IN, KS, KY, MN, MO, NC, NM, NY, OK, PA, SD, TX, and WI. The countries of Algeria, Australia, Belarus, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Cyprus, Czechia, Egypt, France, Germany, Guatemala, Hungary, India, Ireland, Italy, Japan, Kazakhstan, Korea, Republic of, Latvia, Malaysia, Morocco, Myanmar, Nepal, Nigeria, Pakistan, Poland, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, South Africa, Taiwan, Province of China, Turkey, United Arab Emirates, and Viet Nam.

Class IIWhat this means

Use of the product may cause temporary or medically reversible adverse health consequences.

Sources & references

Taylor Coteus
Legal News Analyst

Taylor Coteus is RecallWatch's Legal News Analyst, covering FDA medical-device enforcement and patient-safety litigation. He reviews every recall against the FDA's official openFDA record before it is published and is responsible for corrections and editorial accuracy.

More Class II recalls

Class IIOngoingZ-3334-2026

Wright Medical Technology, Inc. recalls CHARLOTTE CLAW Screw

Screws, intended for use with polyaxial compression plating system intended for foot osteotomies, was manufactured with missing threads on the screwheads, which if used would prevent a secure rigid engagement with the pl…

  • Bone Fixation Plate
  • Process control
Wright Medical Technol…TNOct 7, 2026