ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM)
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001).…
Drawn from openFDA device enforcement reports. Filter by severity, status, state, year, or manufacturer.
Showing 1,465–1,476 of 3,866 recalls
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001).…
For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes.…
Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjust…
When the device is activated during surgery, the device software issues an error message "Instrument Face Mask (6001385000000) is not valid for this application" resulting in an inability to use the device.
Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to f…
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
Abbott Laboratories is recalling their Alinity ci-series System Control Module, a chemistry/immunoassay anayzer, by correction.…
Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may cause inaccurate results.
Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may cause inaccurate results.
When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do…
Removal of affected lot of screws due to labeling error.
Due to firm mistakenly distributing product within the U.S. without FDA premarket clearance/approval.